SPECIFICATIONS
1. Product name: Blood Pressure Monitor
2. Model: KD-558
3. Classification: Internally powered, Type BF applied part, IPX0, No AP
or APG, Continuous operation
4. Machine size: Approx. 138mm × 98mm × 48mm
(5 7/16” x 3 27/32” x 1 7/8”)
5. Cuff circumference: 22cm-30cm (8 21/32”-11 13/16”),
30cm-42cm (11 13/16”-16 17/32”) (Optional),
42cm-48cm (16 17/32”-18 29/32”) (Optional)
6. Weight: Approx. 211g (7 7/16 oz.) (exclude batteries and cuff)
7. Measuring method: Oscillometric method, automatic inflation
and measurement
8. Memory volume: 2x60 times with time and date stamp
9. Power source: batteries: 4 ×1.5V
SIZE AA
10. Measurement range:
Cuff pressure:
0-300 mmHg
Systolic:
60-260 mmHg
Diastolic:
40-199 mmHg
Pulse rate:
40-180 beats/minute
11. Accuracy:
Pressure:
±3 mmHg
Pulse rate:
±5%
12. Environmental temperature for operation: 10°C~40°C (50°F~104°F)
13. Environmental humidity for operation: ≤90%RH
14. Environmental temperature for storage and transport: -20°C~55°C
(-4°F~131°F)
15. Environmental humidity for storage and transport: ≤90%RH
16. Environmental pressure: 80KPa-105KPa
17. Battery life: Approx 500 times
18. A list of all components belonging to the pressure measuring system,
including accessories: Pump,Valve, LCD, Cuff, Sensor
Note: These specifications are subject to change without notice.
NOTICE
1. Read all of the information in the operation guide and any other literature
in the box before operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same arm for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements
to allow the blood circulation in your arm to recover. Prolonged
ENGLISH
Содержание KD-558
Страница 126: ...126 3 1 1 5 START LCD...
Страница 129: ...129 7 1 2 3 4 5 8 9 10 American National Standards Institute 11 12 IHB IHB 1 25 2 0 14s 53 13 14...
Страница 130: ...130 15 16 15 FCC TV 17 ANSI AAMI SP 10 2002 A1 2003 A2 2006 1 a b c LCD...
Страница 131: ...131 2 a LCD 2 2 1 2 2 1 2 2 b START MEM 2 2 START MEM MEM c START d 1 e 3...
Страница 132: ...132 4 a velcro b 1 2 cm c d 1 2 3 4 5 5 200...
Страница 133: ...133 5 a b c a b c 6 a START 6 b U1 U2 6 1 MEM 6 2 START 5 6 6 1 6 2 c 6 3 d LCD...
Страница 134: ...134 6 4 6 5 6 3 6 4 6 5 e 1 START f START 7 a MEM LCD 7 7 7 1 7 2 b MEM 7 1 START 7 2 MEM 5 c LCD...
Страница 135: ...135 7 3 7 4 LCD 7 5 7 3 7 4 7 5 d MEM 7 6 7 7 7 8 MEM 7 9 MEM 7 6 7 7 7 8 7 9 e 1 START...
Страница 137: ...137 10 monitor display LCD HI Lo IEC 60601 1 8 reset display LCD 8 11 1 LCD...
Страница 139: ...139 1 2 3 4 5 6 2 7 8 9 10 000 1000 10 2 75 90 BF...
Страница 140: ...140 93 42 SN GIMA 12 GIMA GIMA software GIMA 0197...
Страница 141: ...141 ARABIC 142 142 143 143 143 143 144 145 146 146 146 147 147 148 149 15 151 151 152 153 154 154 10 11 12...
Страница 143: ...143 ARABIC...
Страница 144: ...144 ARABIC...
Страница 145: ...145 ARABIC...
Страница 146: ...146 ARABIC...
Страница 147: ...147 ARABIC...
Страница 148: ...148 ARABIC...
Страница 149: ...149 ARABIC...
Страница 150: ...150 ARABIC...
Страница 152: ...152 ARABIC 11...
Страница 153: ...153 ARABIC 12...
Страница 154: ...154 ARABIC...
Страница 155: ...155 ARABIC...
Страница 156: ......