15
the
relative
humidity
should
be at
least 30 %.
Electrical
fast
transient/bur
st
IEC
61000-4-4
± 2 kV for
power
supply lines
± 1 kV for
input/output
lines
± 2 kV for
power
supply
lines
± 1 kV for
input/outp
ut lines
Mains
power
quality should be
that of a typical
commercial
or
hospital
environment.
Surge
IEC
61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
±
1
kV
line(s) to
line(s)
±
2
kV
line(s) to
earth
Voltage
dips, short
Interruptions
and Voltage
variations on
power
supply input
lines
IEC
61000-4-11
<5 % UT
(>95 % dip in
UT)
for 0.5 cycle
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 s
<5 % UT
(>95 % dip
in UT)
for
0.5
cycle
40 % UT
(60 % dip
in UT)
for
5
cycles
70 % UT
(30 % dip
in UT)
for
25
cycles
<5 % UT
(>95 % dip
in UT)
for 5 s
Mains
power
quality should be
that of a typical
commercial
or
hospital
environment.
If
the user of our
product requires
continued
operation during
power
mains
interruptions, it is
recommended
that our product
be powered from
an uninterruptible
power supply or a
battery.