Guidance and Declaration - Electromagnetic Emissions
The Blood pressure monitor is intended for use in the
electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an
environment.
Emission tests
Complian
ce
Electromagnetic environment
- guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only
for
its
internal
function.
Therefore, its RF emissions are
very low and are not likely to
cause
any
interference
in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in
all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage
Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
Guidance and Declaration - Electromagnetic Immunity
The Blood pressure monitor is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Blood pressure monitor should assure that it is used in
such an environment.
Immunity
test
IEC60601 test
level
Complian
ce level
Electromagnetic
environment-
guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
±
6
kV
contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered
with
synthetic material,