25
ENGLISH
a) The monitor's window:
Use cotton swab or cloth strip wetted by clean water to graze and remove blot and
naturally air dried. Be sure it is dry before usage.
b) Cleaning and disinfecting of adapter:
Disinfection:
● For single use airway adapter (A3,A3N):
Reuse of the single use Adapter can cause cross infection.
● For reusable airway adapter (rA3)
Soak the adapter in disinfectant and dry it in a ventilated place for at least 30min
after taking it out.
See the following table for specific disinfectant and soaking time requirements:
Disinfectant
Concentration
Soaking time
Chlorine disinfectant
500-1000ppm
30 min
glutaraldehyde
2.0%
20-90min
peracetic acid
0.2%
10min
Used and discarded adapters should be treated according to State regulation.
Be sure it is dry before its usage.