In this menu, to press button S to move the cursor to
choose item, to press button
+/
or button
to change data highlighted by the
cursor.
Some items in this menu, if there are some actions not for changing data but just
for direct operations such as
LOAD_DEFAULTS or EXIT, then to press button
+/
to execute.
In this menu, to press
button, then to exit this menu return back to the main
display screen.
This menu includes the following setups:
1). The high alarm limits of EtCO2: EtCO2 ALARM_H: 22- 99mmHg, off
2). The low alarm limits of EtCO2: EtCO2 ALARM_L: off, 10-60mmHg
3). The high alarm limits of respiration rate: RESP ALARM_H:5-60t/m, off
4). The low alarm limits of respiration rate: RESP ALARM_L: off, 4-40t/m
5). The setup of no CO2 detected time: APNEA TIME: 15s-44s, off
6). The unit of CO2: CO2 UNITÀ %, mmHg or kPA 7). Sensor zero Calibration
8). CO2 Wave scale: WAVE SCALE: 54mmHG or 76mmHG
9). Default reload: LOAD-DEFAULTS
10). Exit: EXIT
Attention:
a) When respiration wave comes and EtCO2 is not zero value, the zero instructive
item ’ZERO’ will be dark colour and zero calibration operation cannot be run;
Only when the sensor is in clean air without respiration wave and EtCO2 is zero
value, can enter the sensor zero calibration item ‘ZERO’, press the
+/
button
then, the sensor can be zero calibrated, but must be sure without breathing near
the sensor during zero calibration
b) The wave scale means the maximum value of waveform amplitude display but it
does not mean data on full-scale. Data on full-scale still means 99mmHg.
Default values as follow:
EtCO2 alarm high limit: 50 mmHg
EtCO2 alarm low limit: 19 mmHg
RESP alarm high limit: 30 rpm
RESP alarm low limit: 08 rpm
APNEA time: 30S
CO2 unit: %
WAVE SCALE: 54mmHg