PERIODICAL MAINTENANCE CHECKS

The TOBI SUCTION ASPIRATOR does not need maintenance or lubrication. 

It is, however, necessary to inspect the unit before each use. With regard to training, given the information 

contained in the user manual and since it is easy to understand the said device, it doesn’t appear to be nec-

essary. 

Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been 

damaged during previous use. Connect the cable to electrical network and turn the switch on. 

Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum 

indicators reaches at least -75kPa (-0.75 bar). Rotate the knob from right to left. The vacuum indicator should 

go down -25kPa (-0.25 bar). 

Check that no loud noises are present. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and 

situated in the plug protects the instrument. For fuses replacing, always the type and the range. 

Before changing the fuse, disconnect the plug from the power supply socket.

If the overfill security system it’s activated, don’t proceede with the liquid aspiration. 

If the overfill security system doesn’t work there are two cases:

1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter 

who avoid the liquid penetration inside the device. 

2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, 

in this case return the device to GIMA technical service. 

Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting in-

structions and any other information that can help the technical assistance staff for product repair.

 BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, 

PLEASE  CONTACT  GIMA  TECHNICAL  SERVICE.  GIMA  DOES  NOT  GIVE  GUARANTEE  IF  INSTRU-

MENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED. 

ENGLISH

Fault type

Cause

Solution

1. The suction unit doesn’t 

work

Cable is damaged External power 

source failure

Replace the cable Check the external power 

source

2. No aspiration

Jar Cap not screwed on correctly

Unscrew the cap, and re-screw it

3. No aspiration

Lid seal not in its seat

Unscrew the cap and insert the seal properly 

in its seat

4. The Vacuum power on the 

patient side is either very low 

or absent

•  Vacuum regulator set 

  to minimum

•  Protection filter blocked or damaged

•  Connection tubes blocked, kinked 

or disconnected

•  Shut-off valve blocked or 

 damaged

•  Pump motor damaged

•  Turn the vacuum regulator clockwise and 

check the value of the vacuum on the gauge

•  Replace the filter

•  Replace or reconnect the tubes, check the jar 

connections 

•  Empty the jar, or disconnect the tube from the 

jar and unblock the shut-off valve. The unit 

twill only work in the upright position 

•  Refer to authorised service personnel

5. The float doesn’t close

If the cap has been washed, ensure 

that the float is not partially detached

Fit the float into it’s place

6. The float doesn’t close

The float it’s covered by dirty material Unscrewed the cap, leave the and put in on 

autoclave

7. Low suction

Foam inside the jar

Fill the jar to 1/3 full of ordinary water

8. No aspiration due to flow 

leakage of mucus  

Filter blocked  

Replace filter  

Faults 1 - 2 - 3 - 4 - 5 - 6 

- 7 - 8

None of the procedures have 

achieved the desired results

Contact GIMA customer service

Содержание 28222

Страница 1: ...HIRURGISCHER ABSAUGER TOBI ASPIRADOR QUIR RGICO TOBI APARELHO DE SUC O CIR RGICO TOBI M28222 M Rev 9 01 21 PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly...

Страница 2: ...fornito dal costruttore GIMA al fine di ga rantire la massima efficienza e sicurezza del dispositivo Non immergere mai l apparecchio in acqua o in altri liquidi Non posizionare o conservare l aspirato...

Страница 3: ...medico viene in contatto con il paziente attraverso la sonda monouso in dotazione con l ap parecchio dotata di relativa certificazione CE conformi ai requisiti della norma ISO 10993 1 per tanto non s...

Страница 4: ...r non entrare in contatto con eventuali sostanze contaminanti dopo ogni ciclo di utilizzo della macchina ACCESSORI IN DOTAZIONE Vaso aspirazione completo 1000ml Sonda aspirazione monouso CH20 Raccordo...

Страница 5: ...ettamente legato all appli cazione dello stesso componente Dopo ogni ciclo di pulizia spetta all utilizzatore finale verifica l idoneit del raccordo al riutilizzo Il componente dovr essere sostituito...

Страница 6: ...ttavia effettuare alcuni semplici controlli per la verifica della funzionalit e della sicu rezza dell apparecchio prima di ogni utilizzo Per quanto riguarda il training visto le informazioni contenute...

Страница 7: ...dell appa recchio ATTENZIONE Per un corretto utilizzo posizionare l aspiratore su una superficie piana e stabile cos da avere il completo volume di utilizzo del vaso e la maggiore efficacia del dispos...

Страница 8: ...ma all interno del vaso di raccolta svitare il coperchio dal vaso e riempire quest ultimo per 1 3 di acqua per facilitare le operazioni di pulizia e velocizzare la depressione durante il funzionamento...

Страница 9: ...nti L uso di accessori trasduttori e cavi diversi da quelli specificati con l eccezione di trasduttori e cavi venduti dal costruttore dell apparecchio e del sistema come parti sostitutive pu risultare...

Страница 10: ...azione dovrebbe essere quel la tipica di un ambiente commerciale o ospedale Se l utente dell ASPIRATORE TOBI richiede che l apparecchio operi in continuazione si raccomanda di utilizzarlo sotto un gru...

Страница 11: ...come un diver so orientamento o posizione dell apparecchio b L intensit di campo su un intervallo di frequenze da 150 kHz a 80 MHz dovrebbe essere minore di 3 V m Distanze di separazione raccomandate...

Страница 12: ...se II Numero di serie Limite di temperatura Fusibile Limite di pressione atmosferica IP21 Grado di protezione dell involucro Corrente alternata Acceso Hz Frequenza di rete Spento Limite di umidit SIMB...

Страница 13: ...ment Do not attempt to make the device work before it has been thorough ly checked by qualified personnel and or the GIMA technical service department Position the device on stable and flat surfaces i...

Страница 14: ...they contain small parts that may be swallowed 14 Do not leave the device unattended in places accessible to children and or persons not in full possession of mental facultiesas they may strangle the...

Страница 15: ...on pump is used on patients in unknown pathological situations and where an assess ment of indirect contamination is not possible The filter is not manufactured to be decontaminated dismantled and or...

Страница 16: ...of local laws and regulations The jug and lid can be further disinfected using a common disinfectant strictly following the instructions and dilution values provided by the manufacturer At the end of...

Страница 17: ...nents list descriptions setting in structions and any other information that can help the technical assistance staff for product repair BEFORE EVERY CHECKING OPERATION IN CASE OF ANOMALIES OR BAD FUNC...

Страница 18: ...the presence of liquids or other visible contamination secretions In the presence of liquids or other visible contamination immediately replace the medical device due to the risk of an insufficient v...

Страница 19: ...d cables differing from those specified with the exception of transducers and cables sold by the appliance and system manufacturer as spare parts can lead to an increase in emissions or in a decrease...

Страница 20: ...ains power quality should be that of a typi cal commercial environment or hospital If the user of the TOBI SUCTION UNIT request that the appliance operates continuously the use of a continuity unit is...

Страница 21: ...es such as changing the device s direction or positioning may be necessary b The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V m Recommended separation distance...

Страница 22: ...EEE disposal Class II applied Serial number Temperature limit Fuse Atmospheric Pressure limit IP21 Covering Protection rate Alternating current ON Hz Mains frequency OFF Humidity limit SYMBOLS Disposa...

Страница 23: ...osants originaux fournis par le fabricant afin de garantir le maximum d efficacit et de s curit du dispositif Utiliser le dispositif avec le filtre de protection originaux fournis par le fabricant afi...

Страница 24: ...aspirateur ne pas alt rer les parties lectriques et ou m caniques S adresser toujours au service technique 11 Le produit et ses pi ces sont bio compatibles conform ment aux exigences de la norme EN 60...

Страница 25: ...uelles substances contaminants apr s claque cycle d usage de l appareil ACCESSOIRES DE SERIE Bocal aspiration complete 1000ml Raccord conique Jeu de tubes 6 mm x 10mm Sonde aspiration CH20 Filtre anti...

Страница 26: ...ycle de st rilisation et le nombre de cycles de nettoyage est strictement li l ap plication de ce composant Par cons quent apr s chaque cycle de nettoyage l utilisateur final doit v rifier la conformi...

Страница 27: ...ntes Let tubes d aspiration en silicone transparent peuvent tre ins r s en autoclave o ils peuvent effectuer un cycle de st rilisation une temp rature de 121 C pression relative 1 bar 15min Le raccord...

Страница 28: ...correctement le cou vercle 2 L appareil n aspire pas Joint du couvercle pas en place D visser le couvercle et remettre le joint en place 3 Le flotteur bloqu Pr sence des incrustations sur le flot teur...

Страница 29: ...ical en raison du risque d un d bit insuffisant de vide Ces produits sont con us test s et fabriqu s exclusivement pour un usage unique et pour une p riode n exc dant pas 24 heures Brancher le ten sil...

Страница 30: ...e transducteurs et de fils autres que ceux qui sont sp cifi s l exception des transducteurs et des fils vendus par le constructeur de l appareil et du syst me comme pi ces de rechange peut provoquer u...

Страница 31: ...5 sec L alimentation devrait tre celle typique d un environnement commercial ou hos pitalier Si l utilisateur de l aspirateur AS PIRATEUR TOBI demande que l appareil op re continuellement il faut uti...

Страница 32: ...ne variation de l orientation ou du positionnement de l appareil b L intensit du champ sur un intervalle de fr quence de 150 kHz 80 MHz doit tre inf rieure 3 V m FRAN AIS Distance de s paration recomm...

Страница 33: ...il de classe II Num ro de s rie Limite de temp rature Fusible Limite de pression atmosph rique IP21 Degr de protection de l enveloppe Courant alternatif Allum Hz Fr quence du secteur Eteint Limite d h...

Страница 34: ...wendet werden Das Ger t auf einer ebenen und stabilen Fl che aufstellen die Luftschlitze d rfen nicht durch Gegen st nde verstopft werden Das Ger t nicht in R umen benutzen in denen brennbare An sthes...

Страница 35: ...sind biokompatibel in bereinstimmung mit den Anforderungen der Norm EN 60601 1 10 Die Funktionsweise des Ger tes ist sehr einfach daher sind die Angaben die im vorliegenden Handbuch geliefert werden v...

Страница 36: ...enn n tig auch Schutz brille und Schutzmaske um nicht mit sch dlichen Substanzen in Ber hrung zu kommen nach jedem Be nutzungsgang SERIENM SSIGES ZUBEH R Flasche saugdruck komplett 1000ml Konusanschlu...

Страница 37: ...t der An wendung des Schlauchs selbst zusammen Daher muss der Endverbraucher nach jedem Reinigungszyklus die Eignung f r einen Wiedergebrauch des Schlauchs berpr fen Das Bauteil muss dann ausgewechsel...

Страница 38: ...Temperatur von 121 C 1 bar Druck 15min autoklaviert werden DEN ANTIBAKTERIELLEN FILTER NIE WASCHEN STERILISIEREN ODER AUTOKLAVIEREN LAUFENDE WARTUNG Das Ger t CHIRURGISCHER ABSAUGER TOBI hat keine wa...

Страница 39: ...em K rper zuvermeiden DEUTSCH Fehler Ursache Abhilfe 1 Keine Absaugung Beh lterstopfen falsch aufgeschraubt Beh lterstopfen abschrauben und wieder richtig aufschrauben 2 Keine Absaugung Stopfendichtun...

Страница 40: ...ellt um bei einem einzigen Patien ten und mit einer maximalen Anwendungsdauer von 24 Stunden verwendet zu werden Verbinden Sie den l ngeren Silikonschlauch mit dem Auslass des Patientenbeh lters Am fr...

Страница 41: ...die Abwesenheit von Anomalien und St rungen si chergestellt wird Der Gebrauch von Zubeh r Wandlern und Kabeln die von den empfohlenen abweichen kann zu einem Zuwachs der Emissionen bzw zu einer Vermi...

Страница 42: ...5 Sek Die Stromeinspeisung muss die f r eine gewerbliche bzw eine Klinikumgebung bliche sein Wenn der Benutzer des SAUGER TOBI verlangt dass das Ge r t im Dauerbetrieb arbeiten soll wird es empfohlen...

Страница 43: ...en n tig werden wie etwa eine andere Ausrichtung oder Stellung des Ger ts b Die Feldintensit t auf einem Frequenzintervall von 150 kHz bis 80 MHz k nnte unter 3 V m liegen Empfohlene Schutztrennabst n...

Страница 44: ...wert Sicherung Luftdruck Grenzwert IP21 Deckungsschutzrate Wechselstrom Ger t Ein Hz Netzfrequenz Ger t Aus Feuchtigkeitsgrenzwert SYMBOLOGIE DEUTSCH Entsorgung Das Produkt darf nicht mit dem anderen...

Страница 45: ...os bacteriologicos prove dos por el fabricante para garantizar la mejora eficacia y seguridad del dispositivo Nunca sumergir el aparato en agua Posicionar el aparato sobre superficies planas y estable...

Страница 46: ...quisitos de la norma EN 60601 1 10 El funcionamiento del dispositivo es muy sencillo y por ello no se requieren ulteriores medidas adem s de lo ya se alado en este manual de uso 11 Ning n de las parte...

Страница 47: ...sustancias contaminantes despu s de cada uso de la m quina ACCESSORIOS DE SERIE Vaso aspiracion completo 1000ml Uni n conica Set tubos 6mm x 10mm Sonda aspiraci n CH20 Filtro antibact rico idrofobico...

Страница 48: ...antidad de ciclos de esterilizaci n y de limpieza est estrictamente relacionado con la aplicaci n del componente en s Por consiguiente tras cada ciclo de limpieza el utilizador final tiene que comprob...

Страница 49: ...NO LAVAR ESTERILIZAR NI TRATAR EN AUTOCLAVE EN NINGUN CASO EL FILTRO ANTI BACTERIANO CONTROL PERI DICO DE MANTENIMENTO El ASPIRADOR QUIR RGICO TOBI no tiene piezas que necesiten mantenimiento y o lub...

Страница 50: ...s con la estructura del aparato ESPA OL Defecto tipo Causa Soluci n 1 El elemento de succi n no funciona Est da ado el cable Apag n de ali mentaci de electricidad del exterior Reponer el cable Comprob...

Страница 51: ...de silicona en la abertura de la tapa del vaso que queda libre En la extremidad que queda libre del tubo largo de silicona conectar la uni n c nica para la conexi n de sondas y por ltimo la sonda de...

Страница 52: ...pecificados con la excepci n de los transductores y cables vendidos por el fabricante del aparato y del sistema como piezas sustituibles puede resultar en un incremento de las emisiones o en una dismi...

Страница 53: ...pica de un entorno comercial o de un hospital Si el usuario del ASPIRADOR TOBI necesita que el aparato funcione continuamente se reco mienda de utilizzarlo bajo un grupo de con tinuidad Campo magn tic...

Страница 54: ...s como una orientaci n o ubicaci n distinta del aparato b La intensidad de campo en un intervalo de frecuencias de 150 kHz a 80 MHz deber a ser menor a 3 V m ESPA OL Distancias de separaci n recomenda...

Страница 55: ...mero de serie L mite de temperatura Fusible L mite de presi n atmosf rica IP21 Tasa de protecci n de cobertura Corriente alterna Encendido Hz Frecuencia de red Apagado L mite de humedad SIMBOLOG A El...

Страница 56: ...so de cair acidentalmente n o tente remover o dispositivo da gua enquanto a ficha estiver ligada desligue o interruptor retire a ficha da tomada el trica e entre em contato com o departamento de assis...

Страница 57: ...uisitos da ISO 10993 1 portanto rea es al rgicas e irrita o da pele n o podem ocorrer 11 O produto e suas pe as s o biocompat veis de acordo com os requisitos da norma EN 60601 1 12 O funcionamento do...

Страница 58: ...acteriano Diferentes vers es dispon veis mediante pedido com reservat rio complete de 2000ml O filtro fabricado com material hidrof bico PTFE para evitar que os fluidos entrem no circuito pneu m tico...

Страница 59: ...ou esterilizar os acess rios A lavagem e ou limpeza do dep sito autoclav vel deve ser realizada como se segue Use luvas de prote o e avental se necess rio culos e m scara de prote o para evitar o con...

Страница 60: ...egula o de aspira o O indicador de aspira o dever descer at 25 kPa 0 25 bar A exist ncia de sons muito ruidosos poder significar maufuncionamento do dispositivo Cont m um fus vel de seguran a F 1x1 6A...

Страница 61: ...nto e ou danos provocados pelo transporte e ou armazenamento A posi o de trabalho deve permitir que seja alcan ado o painel de comandos e que estejam bem vis veis o indicador de v cuo o reservat rio e...

Страница 62: ...ade adicional de desconex o da rede el trica N UTILISER JAMAIS LE DISPOSITIF SANS FLACON ET OU SANS LE FILTRE DE PROTEC TION EVITER QUE DES ENFANTS ET OU DES INCAPABLES PUISSENT UTILISER L APPAREIL SA...

Страница 63: ...magn tico Guia Descarga eletrost tica ESD IEC EN 61000 4 2 8kV contato 15kV ar O dispositivo n o muda de estado O ch o dever ser de madeira cimento ou azulejo Se o ch o se encontrar coberto com um mat...

Страница 64: ...do eletromagn tico da instala o a poder ser menor do que o n vel de conformidade para cada intervalo de frequ ncia b poss vel verificar se h interfer ncia na proximidade de dispositivos identifi cados...

Страница 65: ...a de sa da n o est indicada acima a dist ncia de separa o em me tros m recomendada pode ser calculada utilizando a equa o aplic vel frequ ncia do transmissor onde P pot ncia nominal m xima de sa da do...

Страница 66: ...s detritos dom sticos Os utiliza dores devem levar os aparelhos a serem eliminados junto do pontos de recolha inicados para a re ciclagem dos aparelhos el tricos e eletr nicos CONDI ES DE GARANTIA GIM...

Страница 67: ......

Страница 68: ......

Отзывы: