19
INSTRUCTIONS
Operation using with AC/DC switching adapter
•
Place the unit on a flat, horizontal surface.
•
Connect the short silicon tube
,
with antibacterial filter, to the suction connector. The other tube, with one end connected to the
antibacterial filter with the other end to jar’s lid connector where has been fixed the red float. When the 90% of the volume of the jar
is reached there is the activation of the security float (the float close the aspiration connector on the jar) to the avoid liquid
penetration inside the device.
•
Connect the long silicon tube to the other jar's lid connector
•
Connect the other end of the long silicon tube to the probe plastic connector
then connect the suction probe to it.
•
Connect the switching adapter to the device with the appropriate connector and insert the power cable plug to the power socket.
To start the treatment press the
I
switch to turn it on
•
Set the desired vaccum value (Bar / kPa) with the appropriate vacuum regulator. Turning the handle clockwise increase the vacuum
value: these values can be read on the “vacuum indicator” instrument.
•
To suspend and / o r terminate the treatment, press the switch againg and pull the plug out from the power socket.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the
functionally vacuum) then rescrew the lid on the jar correctly.
•
To extract the accessories and start with cleaning.
WARNING:
Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the “VACUUM”.
A wrong connection causes immediate destruction in case of contact with sucked liquids.
Filter assembling
FLOW DIRECTION
IN
If the state of the protective conductor of the installation, to which the medical device in question is connected, appear to be dubious, the
use of equipment with a internal battery is recommended.
Operation using cigarette lighter DC 12V
•
Connect the device’s external plug 12V to the ligther plug with the cigarette lighter cable. Check the battery power status of the
vehicle before the cigarette lighter cable. Press the switch to start suction
•
Press the switch to the
I
position to turn it on.
WARNING
:
Only use the originally supplied or recommended replacement cigarette ligther cables (view the chapter “Important
Safety Rules”)
WARNING:
Before using the device, check the battery power status. Before each use proceed with charging the battery.
To maintain the device in good conditions, recharge the battery every 3 months (when not in use).
Recharging operations:
to be able to charge the internal battery it is necessary to connect the universal switching adapter
(mod. UE 60140429SPA3 of FUHUA) to the electric network for approx. 120 to 150 minutes with the main switch to position 0.
The battery’s autonomy when fully charged is approx. 60 minutes with continued operations.
IN / Fluid Inside
Suction Pump Inlet
Jar Air Tube