![Gilson VALVEMATE Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/gilson/valvemate/valvemate_user-manual_2227573032.webp)
2-19
Installation
2
V
alve Inst
allation or Replacement
Installing the Replacement Valve
To install the replacement valve:
1 Insert the valve with the adapter’s groove
facing upward. This is the only position in
which the VALVEMATE’s coupler will accept
the valve.
If necessary, loosen the
coupling screws so the shaft
can slide into the coupling.
The coupling screws are
accessed via the two access
holes in the right side of the
VALVEMATE.
2 Take turns tightening the
internal coupling screws via
the access holes in the right
side of the VALVEMATE.
3 Insert the handle of the 7/16"
Allen (hex) wrench into the hole in the
thumbscrew and turn the thumbscrew to the
right.
4 ENSURE THAT ALL TOOLS HAVE BEEN
REMOVED FROM THE VALVEMATE.
5 Turn off then turn on power to the
VALVEMATE.
6 Verify that the display correctly identifies the
installed valve.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание VALVEMATE
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 57: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 65: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 71: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...