QUICK GUIDE 

|

 

PIPETMAN® G

Aspirate

1. 

Press the push button to the 

first stop

 

(this corresponds to the set volume of liquid).

2.  Hold the pipette vertically and immerse the tip 

in the liquid (refer to 

PIPETMAN G User’s Guide 

LT801122).

3.  Release the push button slowly and smoothly (to 

the 

top

 position) to aspirate the set volume of 

liquid.

4.  Wait one second, and then withdraw the pipette 

tip from the liquid. You may wipe any droplets 

away from the outside of the tip using a medical 

wipe; however if you do so, take care to avoid 

touching the tip’s orifice.

First stop

Second stop

Dispense

1. 

Place the end of the tip against the inside wall of the 

recipient vessel (at an angle of 10° to 40°).

2.  Press the push button slowly and smoothly to the 

first stop

.

3.  Wait for at least one second, then press the push 

button to the

 second stop

 to expel any residual 

liquid from the tip. Keep the push button pressed 

fully down and (while removing the pipette) draw 

the tip along the inside surface of the vessel.

4.  Release the push button smoothly. Eject the tip by 

pressing firmly on the tip ejector button.

It is recommended to regularly verify the volume 

displayed on the dial while pipetting.

NOTICE

Figure 1

Pipetting motion - aspirate and dispense

Rest position

Содержание PIPETMAN G

Страница 1: ...Quick Guide EN G...

Страница 2: ...y refer to Fitting the Tips on page 6 Certi cate of Conformity including barcode sticker Multichannel models PIPETMAN G PIPETMAN G Quick Guide Safety bag Ejector spacers for D10 tips for Px10G models...

Страница 3: ...and precision The data provided in the Maximum Permissible Errors table on page 5 were obtained using PIPETMAN DIAMOND Tips These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used...

Страница 4: ...SYSTEMA TIC ERROR L RANDOM ERROR L PIPETMAN G SINGLE CHANNEL P2G P N F144054M WITH D10 OR DL10 TIPS 0 2 0 5 1 2 0 024 0 025 0 025 0 030 0 012 0 012 0 012 0 014 0 08 0 08 0 08 0 08 0 04 0 04 0 04 0 04...

Страница 5: ...e setting slowly reach the required setting making sure to not overshoot the mark When increasing the volume setting pass the required value by 1 3 of a turn and then slowly decrease to reach the volu...

Страница 6: ...122 ACCESSORIES Fitting the Tips on P2G and P10G Models NOTE For the P2G and P10G models a dual position adapter plastic is required to fit DL10 tips long tips or D10 tips short tips The metallic rod...

Страница 7: ...rst volume of liquid and then dispensing it back into the same vessel or to waste Subsequent volumes you pipette will have levels of accuracy and precision within specifications Ejector Spacer for PIP...

Страница 8: ...d touching the tip s orifice First stop Second stop Dispense 1 Place the end of the tip against the inside wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoot...

Страница 9: ...p ejector tip holder seal and O ring NOTICE PIPETMAN P2G and P10G should not be disassembled You may only replace the push button tip ejector and its dual position adapter With these pipettes if the t...

Страница 10: ...or refund of the purchase price paid for such pipettes Routine cleaning control and recalibration are not covered under the warranty The replacement of wearing parts such as seals O rings broken pist...

Страница 11: ...11 PIPETMAN G QUICK GUIDE NOTES...

Страница 12: ...www gilson com contactus LT801624 C 2019 Gilson Inc All rights reserved...

Отзывы: