PIPETMAN® G
|
QUICK GUIDE
QUICK GUIDE
|
PIPETMAN® G
Setting the Volume
The volume of liquid to be aspirated is set using the
volumeter. The dials are colored either black or red to
indicate the position of the decimal point, depending on
the model (refer to Figure 3).
The volume is set by turning
the thumbwheel or the push
button (refer to Figure 4). The
push button makes it easier
and faster to set volumes,
especially when wearing
gloves. The thumbwheel may
be turned using only one hand
to slowly reach the required
setting.
To achieve maximum accuracy
when setting the volume,
proceed as follows:
●
When
decreasing
the
volume setting, slowly
reach the required setting,
making sure to not
overshoot the mark.
●
When
increasing
the
volume setting, pass the
required value by 1/3 of
a turn and then slowly
decrease to reach the
volume, making sure to
not overshoot the mark.
MODEL
COLOR OF VOLUMETER NUMBERS
BLACK
RED
INCREMENT
P2G
µL
0.01 µL
0.002 µL
P10G - P20G - Px10G - Px20G
µL
0.1 µL
0.02 µL
P100G - P200G - Px200G - Px300G
µL
-
0.2 µL
P1000G
0.01 mL
mL
0.002 mL
P5000G
0.01 mL
mL
0.002 mL
P10mLG
mL
0.1 mL
0.02 mL
Figure 3
Dial colors by model
1
5
P2G
1.25 µL
2
0
7
5
P10G
7.5 µL
1
2
5
P20G
12.5 µL
P100G
0
7
5
75 µL
P200G
1
2
5
125 µL
P1000G
0
7
5
0.75 mL
P5000G
1
2
5
1.25 mL
P10mLG
0
7
5
7.5 mL
Push button
Thumbwheel
Figure 4
Location of push button and
thumbwheel
* P5000G and P10mLG do not have tip ejectors.
0
7
5
Px10G
7.5 µL
1
2
5
Px20G
12.5 µL
Px200G
1
2
5
125 µL
Px300G
1
2
5
125 µL