Pound Weight Sets

Model

Total Mass

Total Load at 10:1 

Beam Ratio¹

Included Weight Sets

0.852lb (1/8tsf)

1.704lb (1/4tsf)

3.409lb (1/2tsf)

6.818lb (1tsf)

13.635lb (2tsf)

27.270lb (4tsf)

HMA-727

HMA-732

HMA-729

54.5lb

109.1lb

218.2lb

  545lbf (8tsf)

  1,091lbf (16tsf)

  2,182lbf (32tsf)

2

2

2

1

1

1

1

1

1

1

1

1

1

1

1

1

3

7

¹tsf values indicate force applied to a 2.50in diameter specimen using a 10:1 beam ratio.

Kilogram Weight Sets

Model

Total Mass

Total Load at 10:1 

Beam Ratio

Included Weight Sets

1kg

4kg

8kg

HMA-730

HMA-725

HMA-731

32kg

64kg

88kg

320kg

640kg

880kg

4

4

4

3

5

5

2

5

8

Description

Consolidometer

Calibration 

Disc

50mm (1.969in) Fixed Ring Consolidometer

2in (50.8mm) Fixed Ring Consolidometer

60mm (2.36in) Fixed Ring Consolidometer

2.42in (61.5mm) Fixed Ring Consolidometer

2.5in (63.5mm) Fixed Ring Consolidometer

70mm (2.756in) Fixed Ring Consolidometer

75mm (2.95in) Fixed Ring Consolidometer

3in (76.2mm) Fixed Ring Consolidometer

HMA-83A

HMA-83B

HMA-83C

HMA-83D

HMA-83E

HMA-83F

HMA-83G

HMA-83H

HMA-88A

HMA-88B

HMA-88C

HMA-88D

HMA-88E

HMA-88F

HMA-88G

HMA-88H

50mm (1.97in) Floating Ring Consolidometer

2in (50.8mm) Floating Ring Consolidometer

2.42in (61.47mm) Floating Ring Consolidometer

2.5in (63.5mm) Floating Ring Consolidometer

70mm (2.76in) Floating Ring Consolidometer

HMA-84A

HMA-84B

HMA-84D

HMA-84E

HMA-84F

HMA-88A

HMA-88B

HMA-88D

HMA-88E

HMA-88F

OPTIONAL ACCESSORIES 

(Purchased Separately)

•  HMA-89 Consolidation Load Frame Stand 
•  Calibration Discs (refer to Consolidometer chart above)
•  HMA-608 Consolidation Data Acquisition Software

UNPACKING & SETUP

1.  Inspect the HM-353 for damage, remove it from the pallet.
2.  Ensure the following items are included:

•  Adjustable counterbalance weight with threaded rod and 

nuts (x2)

•  Dial Indicator support rod
•  Dial Indicator adjusting bracket
•  Weight hanger with connecting pin
•  Load-holding screw with adjusting knob
•  Anchor bolts (x2), nuts (x4) and washers (x4)

3.  Unpack the required accessories (ordered separately):

•  Weight Set(s)
•  Consolidometer(s)
•  Dial Indicator or LVDT and Digital Readout Box

4.  Fasten the Load Frame to a sturdy table using the included 

3/8” diameter by 5” long Anchor Mounting Bolts, making sure 

there is enough clearance for the load-holding screw and the 

weight hanger. 

4a.   Fasten the Load Frame to the Load Frame Stand. The stand 

includes a front and back leg with four ¾”-16 x 1” Hex Head 

Bolts, and four Hex Nuts.  
  Remove the two screws on the back underside of the load 

frame base.  Attach the back leg to this area using the longer 

screws that were included with the stand.  
Bolt the font leg into the base using the two of the included 

bolts and two hex head bolts.

OPERATING INSTRUCTIONS

1.  Read all safety and operating instructions before operating 

the unit.

2.  Consult ASTM D2435, ASTM D4546 or AASHTO T 216 for spe-

cific instructions on the testing procedures.

3.  Fasten the weight hanger to the unit and adjust using the 

connecting pin.

4.  Secure the Dial Indicator support rod toward the back of the 

base, fasten the Dial Indicator holder to the rod and then 

connect the Dial Indicator or LVDT to the holder.

5.  Insert the counterbalance weight with threaded rod approxi-

mately 1” and tighten the jam nut.

6.  Balance the lever with the top loading arm by adjusting the 

counterweight and tighten the nut.

7.  Set the Consolidometer (with sample) on the platform and 

turn the adjusting screw until contact is made with the load 

pad in the Consolidometer.

8.  Adjust the Dial Indicator or LVDT to the desired setting.