Description

The pipetting head moves both vertically and horizontally. Vertical movements 

are made to fit tips and pipette liquid. Horizontal movements are made to 

transfer liquids. This movement is possible only when the head is fully raised.
Two drawers can hold up to three different microplates and can be used at the 

same time.

Height Adjuster

The descent of the pipetting head and the 
position of the tips can be adjusted by using  
three height adjusters, as shown in figure 9.

 

.

This will prevent the tips from touching 

the bottom of the microplate and can be 

adjusted for use with shallow- or deep-well 

microplates.

Figure 8

PLATEMASTER® lock

Figure 9

Three height adjusters positions

Tip ejector

Movable drawers

Pipetting head height adjuster

Pin-plate

Moving PLATEMASTER

Whether placed on the lab bench, under a fume hood, 

in a cold room, or used in the field, PLATEMASTER is 
completely portable. PLATEMASTER includes a locking 
system to lock the head when the instrument is being 

moved.
After each use of PLATEMASTER, remember to lock the 

pipetting head to prepare for transport.
To lock the pipetting head, pull the black part on the 

left side of PLATEMASTER, lower the head, and turn 

the black part until the notch is in the slot. Let the head 

slowly ride up and be locked.
To unlock the pipetting head, pull the black part and turn it. Once the pipetting head 

is released, check proper movement of the device.

Never move the device without locking the pipetting head. 

Damages caused by movement without locking the device will not 

be covered by the warranty.

NOTICE

1

1

O

PE
R

A

TIO

N

7

PLATEMASTER®

 

|

 

USER’S GUIDE

Содержание F110761

Страница 1: ...User s Guide EN...

Страница 2: ...iption 7 Moving PLATEMASTER 7 Height Adjuster 7 Volume Setting 8 Tip Fitting 8 Pipetting 9 Pipetting 384 wells 10 Tip Ejection 11 General Guidelines for Good Pipetting 12 ACCESSORIES 13 ASSOCIATED TIP...

Страница 3: ...pipetting steps Figure 1 Gilson PLATEMASTER P20 PLATEMASTER is suitable for a wide range of applications Pipetting and providing master mixes for PCR qPCR Simultaneous processing of nearly all reacti...

Страница 4: ...filter tips installed can be autoclaved Gilson is not liable for any damage resulting from the use of PLATEMASTER Follow the instructions in this user s guide to ensure durability and top performance...

Страница 5: ...vided by the chemical manufacturer before cleaning up the spill and take all safety precautions required Remove PLATEMASTER pipetting head and PLATEMASTER base from box Insert the pipetting head as fo...

Страница 6: ...F110761 F110762 Volume Range 0 5 to 20 L Volume can be adjusted in 0 1 L increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C...

Страница 7: ...ee height adjusters positions Moving PLATEMASTER Whether placed on the lab bench under a fume hood in a cold room or used in the field PLATEMASTER is completely portable PLATEMASTER includes a locking...

Страница 8: ...le Move the pipetting head to the left side Slide the right drawer out and place a new tip rack on the back position The four locations have been specially designed for PIPETMAN DIAMOND Tips TIPACK Sl...

Страница 9: ...ze the pipetting handle to the first stop this corresponds to the set volume of liquid Use the pipetting handle to lower the pipetting head to immerse the tips in the liquid see immersion depth table...

Страница 10: ...pper screw so that the pipetting head may only be moved down to a position slightly above the bottom Pipetting 384 well Microplates PLATEMASTER is able to load 384 well plates with only four movements...

Страница 11: ...Move the pipetting head over the microplate and dispense the last 96 wells The 384 well microplate is loaded in less than one minute requiring only four liquid aspiration steps Tip Ejection Move the...

Страница 12: ...id used Never allow any liquid to enter the pin plate see Figure 14 page 15 This phenomenon can be prevented by squeezing and retracting the pipetting handle slowly and gently Change the tips before a...

Страница 13: ...ad height adjuster F1077605 Lubrication box for O rings PLATEMASTER F1077606 AmpliPur Expert Tips Pin plate Assembly for PLATEMASTER F110765 AmpliPur Expert Tips Reload Block pack of 2 for PLATEMASTER...

Страница 14: ...0ST Validated for use with 384 well plates PIPETMAN DIAMOND Tips DL10 DL10ST DSL10 and DSL10ST could be used under a specific product configuration For more information please contact Gilson Technical...

Страница 15: ...EMASTER to excessive UV radiation and never overnight Do not autoclave any part of PLATEMASTER Material in Contact with Liquids or Vapors FKM Fluoroelastomers EPDM ethylene propylene diene monomer PP...

Страница 16: ...PIPETMAN DIAMOND Tips D300 and AmpliPur Expert Tips Each PLATEMASTER is inspected and validated by qualified technicians in accordance with the Gilson Quality System The adjustment is carried out und...

Страница 17: ...in the event devices are altered or modified by any party other than Gilson or its designates Gilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacem...

Страница 18: ...NOTES Notes 18 Notes PLATEMASTER...

Страница 19: ...NOTES Notes 19 PLATEMASTER USER S GUIDE...

Страница 20: ...e without notification errors excepted Reproduction adaptation or translation without prior written permission is prohibited except as allowed under copyright laws LT801549 I 2022 Gilson Inc All right...

Отзывы: