background image

Chapter 6

SERVICE

GLP Features

The 

serial number

 is engraved on the back of the pipetting head. It provides unique 

identification of your PLATEMASTER and the date of manufacture.

The 

certificate of conformity

 provides traceability of your PLATEMASTER.

Cleaning and Decontamination

Do not use corrosive and acid reagents to clean or decontaminate PLATEMASTER.

PLATEMASTER must be cleaned before it is decontaminated. Wipe the surface of the 

instrument with alcohol or laboratory disinfectant.

PLATEMASTER surface can be cleaned with a soft tissue moistened with water or 

laboratory disinfectant.

Do not expose PLATEMASTER to excessive UV radiation and never overnight. 
Do not autoclave any part of PLATEMASTER.

Material in Contact with Liquids or Vapors

FKM (Fluoroelastomers), EPDM (ethylene propylene diene monomer), 

PP (Polypropylene), POM (Polyoxymethylene), PET (Polyethylene terephthalate), 
Silicone, Stainless steel, Anodized aluminium.

Use of the Lubrication Box

In case of intensive use between each 

maintenance interval and to avoid bad tip fitting, 

it is recommended to lightly lubricate the external 

O-ring using the lubrication box (part number 

F1077606).

Proceed as follows

• 

Open the box lid

• 

Follow the “Tip Fitting” instructions (refer to 

Tip Fitting

 on page 8)

• 

Eject the tips.

The O-ring lubrication is complete. 

Using a no-fiber wipe, remove any excess lubricant on the pin-plate.

Figure 16

Pin-plate

S

E

R

V

IC

E

15

PLATEMASTER®

 

|

 

USER’S GUIDE

Содержание F110761

Страница 1: ...User s Guide EN...

Страница 2: ...iption 7 Moving PLATEMASTER 7 Height Adjuster 7 Volume Setting 8 Tip Fitting 8 Pipetting 9 Pipetting 384 wells 10 Tip Ejection 11 General Guidelines for Good Pipetting 12 ACCESSORIES 13 ASSOCIATED TIP...

Страница 3: ...pipetting steps Figure 1 Gilson PLATEMASTER P20 PLATEMASTER is suitable for a wide range of applications Pipetting and providing master mixes for PCR qPCR Simultaneous processing of nearly all reacti...

Страница 4: ...filter tips installed can be autoclaved Gilson is not liable for any damage resulting from the use of PLATEMASTER Follow the instructions in this user s guide to ensure durability and top performance...

Страница 5: ...vided by the chemical manufacturer before cleaning up the spill and take all safety precautions required Remove PLATEMASTER pipetting head and PLATEMASTER base from box Insert the pipetting head as fo...

Страница 6: ...F110761 F110762 Volume Range 0 5 to 20 L Volume can be adjusted in 0 1 L increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C...

Страница 7: ...ee height adjusters positions Moving PLATEMASTER Whether placed on the lab bench under a fume hood in a cold room or used in the field PLATEMASTER is completely portable PLATEMASTER includes a locking...

Страница 8: ...le Move the pipetting head to the left side Slide the right drawer out and place a new tip rack on the back position The four locations have been specially designed for PIPETMAN DIAMOND Tips TIPACK Sl...

Страница 9: ...ze the pipetting handle to the first stop this corresponds to the set volume of liquid Use the pipetting handle to lower the pipetting head to immerse the tips in the liquid see immersion depth table...

Страница 10: ...pper screw so that the pipetting head may only be moved down to a position slightly above the bottom Pipetting 384 well Microplates PLATEMASTER is able to load 384 well plates with only four movements...

Страница 11: ...Move the pipetting head over the microplate and dispense the last 96 wells The 384 well microplate is loaded in less than one minute requiring only four liquid aspiration steps Tip Ejection Move the...

Страница 12: ...id used Never allow any liquid to enter the pin plate see Figure 14 page 15 This phenomenon can be prevented by squeezing and retracting the pipetting handle slowly and gently Change the tips before a...

Страница 13: ...ad height adjuster F1077605 Lubrication box for O rings PLATEMASTER F1077606 AmpliPur Expert Tips Pin plate Assembly for PLATEMASTER F110765 AmpliPur Expert Tips Reload Block pack of 2 for PLATEMASTER...

Страница 14: ...0ST Validated for use with 384 well plates PIPETMAN DIAMOND Tips DL10 DL10ST DSL10 and DSL10ST could be used under a specific product configuration For more information please contact Gilson Technical...

Страница 15: ...EMASTER to excessive UV radiation and never overnight Do not autoclave any part of PLATEMASTER Material in Contact with Liquids or Vapors FKM Fluoroelastomers EPDM ethylene propylene diene monomer PP...

Страница 16: ...PIPETMAN DIAMOND Tips D300 and AmpliPur Expert Tips Each PLATEMASTER is inspected and validated by qualified technicians in accordance with the Gilson Quality System The adjustment is carried out und...

Страница 17: ...in the event devices are altered or modified by any party other than Gilson or its designates Gilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacem...

Страница 18: ...NOTES Notes 18 Notes PLATEMASTER...

Страница 19: ...NOTES Notes 19 PLATEMASTER USER S GUIDE...

Страница 20: ...e without notification errors excepted Reproduction adaptation or translation without prior written permission is prohibited except as allowed under copyright laws LT801549 I 2022 Gilson Inc All right...

Отзывы: