Chapter 6
SERVICE
GLP Features
The
serial number
is engraved on the back of the pipetting head. It provides unique
identification of your PLATEMASTER and the date of manufacture.
The
certificate of conformity
provides traceability of your PLATEMASTER.
Cleaning and Decontamination
Do not use corrosive and acid reagents to clean or decontaminate PLATEMASTER.
PLATEMASTER must be cleaned before it is decontaminated. Wipe the surface of the
instrument with alcohol or laboratory disinfectant.
PLATEMASTER surface can be cleaned with a soft tissue moistened with water or
laboratory disinfectant.
Do not expose PLATEMASTER to excessive UV radiation and never overnight.
Do not autoclave any part of PLATEMASTER.
Material in Contact with Liquids or Vapors
FKM (Fluoroelastomers), EPDM (ethylene propylene diene monomer),
PP (Polypropylene), POM (Polyoxymethylene), PET (Polyethylene terephthalate),
Silicone, Stainless steel, Anodized aluminium.
Use of the Lubrication Box
In case of intensive use between each
maintenance interval and to avoid bad tip fitting,
it is recommended to lightly lubricate the external
O-ring using the lubrication box (part number
F1077606).
Proceed as follows
•
Open the box lid
•
Follow the “Tip Fitting” instructions (refer to
on page 8)
•
Eject the tips.
The O-ring lubrication is complete.
Using a no-fiber wipe, remove any excess lubricant on the pin-plate.
Figure 16
Pin-plate
S
E
R
V
IC
E
15
PLATEMASTER®
|
USER’S GUIDE