background image

Page 3 of 5 

LT375008-04

GX-271 Liquid Handler Installation Qualification

2. Pre-Installation

Instrument Description

The GX-271 Liquid Handler is an X/Y/Z instrument that automates liquid handling procedures. The liquid handler 

can be configured with a GX Solvent System or VERITY® Syringe Pump, rinse stations, and racks. Optional accessories 

include a GX Direct Injection Module, GX Rinse Pump, and fraction collection valve.

For technical specifications for the GX-271 Liquid Handler, refer to the 

Chapter 1, Introduction

 in the 

GX-271 Liquid Handler User's Guide

. For technical specifications for an optional accessory, refer to its appendix in the 

GX-271 Liquid Handler User's Guide

.

Unpacking

When unpacking the instrument, check the contents against the information listed under the 

Unpacking

 heading in 

Chapter 1, Introduction

 in the 

GX-271 Liquid Handler User's Guide

Retain all packing material so the GX-271 Liquid Handler may be shipped safely in the future.

Installation Site Requirements

Technical Specification

Definition

Power Requirements

50 to 60 Hz

Voltage: 100 to 240V (Universal Input)

Current 

Rating

2A for 100 to 120V

1A for 220 to 240V

Power Consumption: 250W (Maximum)

Temperature Range

5°C to 40°C

Dimensions 

(W x D x H)

59.7 x 54.1 x 57.1 cm (23.5 x 21.3 x 22.5 in.)

NOTE

The height dimension does not include the Z-arm, which will vary depending on 

where the Z-arm is clamped.