Page 3 of 5
LT375008-04
GX-271 Liquid Handler Installation Qualification
2. Pre-Installation
Instrument Description
The GX-271 Liquid Handler is an X/Y/Z instrument that automates liquid handling procedures. The liquid handler
can be configured with a GX Solvent System or VERITY® Syringe Pump, rinse stations, and racks. Optional accessories
include a GX Direct Injection Module, GX Rinse Pump, and fraction collection valve.
For technical specifications for the GX-271 Liquid Handler, refer to the
Chapter 1, Introduction
in the
GX-271 Liquid Handler User's Guide
. For technical specifications for an optional accessory, refer to its appendix in the
GX-271 Liquid Handler User's Guide
.
Unpacking
When unpacking the instrument, check the contents against the information listed under the
Unpacking
heading in
Chapter 1, Introduction
in the
GX-271 Liquid Handler User's Guide
.
Retain all packing material so the GX-271 Liquid Handler may be shipped safely in the future.
Installation Site Requirements
Technical Specification
Definition
Power Requirements
50 to 60 Hz
Voltage: 100 to 240V (Universal Input)
Current
Rating
2A for 100 to 120V
1A for 220 to 240V
Power Consumption: 250W (Maximum)
Temperature Range
5°C to 40°C
Dimensions
(W x D x H)
59.7 x 54.1 x 57.1 cm (23.5 x 21.3 x 22.5 in.)
NOTE
The height dimension does not include the Z-arm, which will vary depending on
where the Z-arm is clamped.