Exceptionally cool or warm ambient temperatures may cause fluctuations in test results. Repeated testing 
or  hot  weather  may  heat  up  the  vessel  body,  and  show  test  results  that  are  artificially  high.  Allow  the 
vessel to cool down before further testing. In cold weather, it may help to run two or three tests in rapid 
succession to warm up the vessel.

The maximum recommended particle size is 0.75in (20mm). Some test methods may have a smaller size 
limit. 

Typical Calibration Curve

Calcium Carbide Tester

W

at

er C

on

ten

t in P

er

cen

t of D

ry M

ass

by D 2216 (

O

ven D

ry)

Gauge Reading, Calcium Carbide Tester

 4.0  ACCURACY

When testing soils, it is recommended that the procedures in ASTM D4944, 

Standard Test Method for 

Field  Determination  of  Water  (Moisture)  Content  of  Soil  by  the  Calcium  Carbide  Gas  Pressure  Tester 
Method

  be  followed.  This  will  ensure  accurate  and  reliable  results.  For  optimum  accuracy  with  most 

materials, the Aqua-Check results should be correlated with oven-dry tests for each material as shown 
in this summary. This simple procedure is outlined in detail in D4944.

1.  Select several samples of the material to be tested having a wide range of water content.
2.  Carefully divide each sample into two representative specimens.
3.  Taking care to not lose moisture during handling, run a test on one specimen in the Aqua-Check, and 

record the gauge reading. Oven dry the other specimen according to the procedure and calculate the 
percent moisture.

4.  Plot the oven-dry moisture on a graph vs. the gauge readings and draw a best-fit curve through the 

points. A wide scatter in the points may indicate operation errors or that the material is not suitable 
for this comparison.

5.  When  testing,  find  the  point  on  the  graph  where  the  gauge  reading  intersects  the  correlation  curve 

and record that value as the corrected percent.

Содержание Aqua-Check MA-26X

Страница 1: ...isture Tester OPERATING MANUAL Rev 5 2022 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 740 548 7298 Email customerservice gilsonco com Product Web Page globalgilson com FAX 800 255 5314...

Страница 2: ...k Moisture Tester MA 26X 2 Safety Instructions 3 1 0 Unpacking Set Up 4 2 0 Introduction 4 3 0 Operating Instructions 5 4 0 Accuracy 8 5 0 Troubleshooting 9 6 0 Accessories 9 Gilson Aqua Check Moistur...

Страница 3: ...t procedures WARNING The calcium carbide reagent used with this test must be handled with great care When exposed to water calcium carbide produces highly flammable or explosive acetylene gas Testing...

Страница 4: ...o order are listed separately in this manual Calcium Carbide Reagent is available separately from Gilson as MAA 44 in a 10lb can 2 0 INTRODUCTION The Gilson Aqua Check Moisture Tester is designed to m...

Страница 5: ...il the balance shows the required 20g NOTE If the moisture content of the material is expected to exceed 20 a 10g sample should be used but the value read on the gauge must then be doubled It is also...

Страница 6: ...mber STEP 6 Holding the Aqua Check Vessel horizontally place the chamber cap on the end of the chamber Position the black dot on the cap so that it will be facing away from your body when the pressure...

Страница 7: ...uld be repeated for 1 3 minutes or until the gauge reading in Step 8 has stabilized b With Pulverizing Balls Hold the Aqua Check horizontally and use an orbital motion to spin the balls around inside...

Страница 8: ...ressure Tester Method be followed This will ensure accurate and reliable results For optimum accuracy with most materials the Aqua Check results should be correlated with oven dry tests for each mater...

Страница 9: ...and clean chamber cap gasket Inspect Vessel cap and stirrup for damage or cracks Replace gasket if necessary X X Defective gauge Gauge needle should rest at zero and move smoothly Replace if defective...

Отзывы: