Gilson Aqua-Check Moisture Tester, MA-26X
6
STEP 5: Adding calcium carbide reagent to
chamber cap
STEP 5.
Using the Long-Handled reagent Scoop, add
at least two full scoops of calcium carbide reagent to
the Chamber Cap.
NOTE:
Make sure you have read and
understand the Material Safety Data Sheet
(MSDS) included with each container of
calcium carbide reagent.
STEP 6: Placing chamber cap on the end of
the chamber
STEP 6: Tightening the top screw to seal
the chamber
STEP 6.
Holding the Aqua-Check Vessel horizontally, place the chamber cap on the end of the chamber.
Position the black dot on the cap so that it will be facing away from your body when the pressure is
released. Move the stirrup into place over the cap and tighten the top screw to seal the chamber. Use
care during this step not to mix the reagent and sample before the chamber is sealed.
STEP 4: Placing Steel Pulverizing
Balls into Vessel
STEP 4: Placing sample into Vessel
STEP 4.
Place the prepared and weighed sample into the Aqua-Check Vessel. For cohesive samples
as described above, also place the two Steel Pulverizing Balls into the Vessel.