NUMBER
O01-1000
REVISION
D
7-8-16
Page
5
of
10
TITLE
TCT-1000 OPERATING INSTRUCTIONS
Teledyne Battery Products
4. CALIBRATION
4.1.
TEST EQUIPMENT
4.1.1. TCT-1000 Battery Capacity Analyzer Test
Fixture: Kelly Aerospace P/N 20758 Test
Fixture (Pictured at right)
4.1.2 Battery: 24 Volt, fully charged per
Manufacturer’s recommendation
4.1.3 Battery: 12 Volt, fully charged per manufacturers recommendation
4.1.4 DC Power Supply: Capable of at lease 24 volts and 20 amps with course and
fine voltage adjustments
4.1.5 Battery Chargers: 4.1.5 (a) or 4.1.5 (b)
a. (1) 12 Volt Charger & (1) 24 Volt Charger
b. (1) At least one DC Power Supply capable of 28 volts or any other means for
charging batteries to manufacturer’s recommendation.
4.1.6 Calibrated FLUKE 87 or Equivalent Multi-meter with
DC volts & milli-volts
ranges (DVM)
4.1.7 Stop watch
4.1.8 Small common screwdriver for adjusting P1 through P4 calibration pots
4.1.9 Miscellaneous hand tools for connecting Test Fixture to batteries and power
supply.
4.1.10 Photocopy of Page 8 showing Records table with blank values in Measured
Value, Pass & Fail columns
5
CALIBRATION PROCEDURE
5.1 Connections
a. Connect 20758 test fixture to 24 volt 20 amp power supply by connecting the red
lead to positive terminal of power supply and the black lead to the negative
terminal of power supply.
b. Plug the connector for the 20758
test fixture into the TCT-1000 connector.
c. Connect the DVM leads to the power supply leads.
d. Record the DVM calibration expiration date in the blank below the Records table.
5.2 20.0 Volt Cutoff
a. It may be necessary to repeat procedure 5.2. & 5.3 until all values fall within
range. When this is necessary there will be interim values that do no need to be
recorded in the Records section. The “final” readings to be recorded are the
readings in which no adjustment is required to complete Procedure 5.3.
b. Set the DVM to DC-Volts and a range capable of measuring up to 22 volts.
c. Turn
Power Switch
off.
d. Adjust power supply until DVM reads 21.0 volts.