Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver 65
Page 58
Issue 4
Doc. No. 1390-PS-0039
June 2016
Do NOT remove black “rubber” transducer caps
.
10.2.
Servicing
There are no moving parts or user-serviceable parts requiring routine maintenance.
Opening the unit or breaking the security seal will void the warranty and the calibration.
In the event of failure, prior to returning the unit to your authorised Gill distributor, it is
recommended that:
All cables and connectors are checked for continuity, bad contacts, corrosion etc.
A bench test is carried out as described in Section 10.6.
Contact your supplier for advice if failure persists.
10.3.
Fault-finding
10.4.
Safe Mode
If a unit is received that will not communicate or the configuration settings are not known
then Safe Mode can be used to establish communication with the WindObserver 65 and
change configuration settings.
Initial Set Up.
The WindObserver 65 RS422 Pair 1 and Pair 2 TX and RX output connections (4 wires)
are connected to a PC via an RS422 to RS232/USB Converter.
Open a Terminal program e.g. Windows HyperTerminal (or use Gill Wind Software as a
Terminal program).
Symptom
Solution
No output
Check DC power to WindObserver 65, cable and connections.
Check comms settings of WindObserver 65 (as detailed in
Section 8) and host system match, including correct Com port.
Check unit is in Continuous mode.
Check that in-line communication devices are wired correctly.
NOTE: It is usual for Anemometer TX + to be connected to
converter device RX +.
Corrupted output
Check comms settings of WindObserver 65 and host system
match.
Try a slower baud rate.
Check cable lengths and type of cable.
One way
communication
Check wiring is in accordance with the manual.
Failed / Incorrect
WindObserver 65
output, data
invalid flag
Check that transducer path is not blocked.