Command Responses
Command responses are only issued when in interactive mode.
Configuration Commands
On receipt of a configuration command, the anemometer will respond with a validation
message if the command was successfully actioned, or an appropriate error message if
it was not. If a configuration command is entered with no data parameter, the current
data value is returned.
Interrogation Commands
On receipt of an interrogation command, the anemometer responds with the requested
information if the command was successfully completed, or an appropriate error
message if it was not.
Invalid Combinations of Commands
It is possible through the setting of various parameters to request a configuration where the required
output bit rate exceeds the configured baud rate.
If an issued configuration command requests a parameter change that would cause the above
situation, an appropriate error message is generated.
COMMANDS
The command descriptions use a modified form of the BNF format:-
|
Option, one option from a list separated by the vertical bar must be included in
the
command.
()
Parenthesis, used to make the possible combinations unambiguous.
[]
Option, the symbols within the square braces are optional.
n{}m
Repeat, the symbols inside the braces can be repeated n to m times. If n is not
included then 0 is implied, if m is not present then infinity is implied.
<>
Non literal symbol, this represents other non literal and literal symbols. A
further definition is required to define this non literal symbol.
IM
Syntax: IM
Description: Switches from measurement mode to interactive mode.
EXIT
Syntax: EXIT
Description: Exits from interactive mode back to measurement mode.
AVERAGE
Syntax:
AVERAGE 2....250
Description: Sets the number of samples used in the block average.
WINDREP
Syntax:
WINDREP ( UVW | POLAR ) ( CAL | UNCAL ) | AXIS
Description: Sets the wind measurement reporting format.
Parameters: UVW Reports wind velocity in UVW format.
POLAR
Reports wind velocity in polar + W format.
AXIS
Reports wind as axis velocities.
CAL
Wind
results
are
calibrated.
HS-50 Research Anemometer
Page 29
Issue 02
Doc. No. 1199-PS-0032
November 2004
Содержание HS-50
Страница 2: ......
Страница 3: ...Gill Instruments Ltd HS 50 Research Anemometer Page 1 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 39: ...Gill Instruments Ltd 11 DRAWINGS HS 50 Research Anemometer Page 37 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 55: ...Gill Instruments Ltd HS 50 Research Anemometer Page 53 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 56: ...Gill Instruments Ltd HS 50 Research Anemometer Page 54 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 57: ...Gill Instruments Ltd HS 50 Research Anemometer Page 55 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 58: ...Gill Instruments Ltd HS 50 Research Anemometer Page 56 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 59: ...Gill Instruments Ltd HS 50 Research Anemometer Page 57 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 60: ...Gill Instruments Ltd HS 50 Research Anemometer Page 58 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 62: ...Gill Instruments Ltd HS 50 Research Anemometer Page 60 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 63: ...Gill Instruments Ltd HS 50 Research Anemometer Page 61 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 64: ...Gill Instruments Ltd HS 50 Research Anemometer Page 62 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 65: ...Gill Instruments Ltd HS 50 Research Anemometer Page 63 Issue 02 Doc No 1199 PS 0032 November 2004 ...
Страница 66: ...Gill Instruments Ltd HS 50 Research Anemometer Page 64 Issue 02 Doc No 1199 PS 0032 November 2004 ...