With the Sensor connected and outputting data to the PCI box.
Check that the unit is correctly configured by going into Configuration mode and
using
D3
(See Page 37).
If a HyperTerminal connection is established to change the sensor configuration
then when a PC keystroke is undertaken then the PCI box Red Tx LED at the top
right on the PCB will be seen to momentarily flash on and off. This indicates a
good connection between the PC and the PCI box.
1.
Check for normal output data, and that the Status Code is OK – 00 (or A for
NMEA format).
2.
If the status code is other than these, refer to Page 41 Status (error) codes.
3.
Use an office fan or similar to check that the unit is sensing wind, turning the unit
to simulate changing wind direction and to check that both axes are functioning.
4.
Note that this is a quick functional test. There are no calibration adjustments; the
unit is designed NOT to require re-calibration within its lifetime.
Use of the Protective Head Cover for an Integrity Check
An Integrity Check is designed to:
1.
Identify any gross changes in the head geometry that would affect the
performance.
2.
Confirm the IS WindObserver zero calibration.
The Integrity Check must be used in an indoor still air environment with an ambient
temperature between 17°C and 23°C. When conducting the test it is important that the
protective cover is assembled on to the IS WindObserver head and not touched or moved
during the test.
Tx LED
Содержание 1086-G-045
Страница 66: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 66 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 67: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 67 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 68: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 68 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 69: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 69 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 71: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 71 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 72: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 72 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 74: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 74 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 75: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 75 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 76: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 76 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 78: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 78 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 79: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 79 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 80: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 80 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 81: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 81 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 83: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 83 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 84: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 84 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 86: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 86 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 87: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 87 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 88: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 88 Issue 10 Doc No 1360 PS 0001 May 2015...
Страница 89: ...Gill Instruments Ltd Intrinsically Safe WindObserver Anemometer Page 89 Issue 10 Doc No 1360 PS 0001 May 2015...