Weight calibration and linearization
During the calibration phase, the display shows the weight intermittently with the word "CAL".
During the linearization phase the display shows the weight intermittently with the wording "LIN n" where in-
stead of n there is the number of the point to be set (from 1 to 9).
Zero calibration
Carry out the operation with the scales unloaded but complete with the tare weight, with stabilized weight. The
displayed weight must be reset. This operation can be repeated several times.
Full scale calibration
Before carrying out the operation, load the sample weight on the scale and wait for it to stabilize; the display
shows the measured value to be calibrated. If the set value is higher than the resolution offered by the instru-
ment, it is not accepted and the display shows an error message for a few seconds. It is always possible to
repeat the calibration operations.
If the instrument is switched off without leaving the set-up menu, the programmed settings will
not be memorized.
!