7 | Operation
G 1501
28 / 37
B-H86.0.12.DB2-2.1
3.
By briefly pressing the
Function key
, you can scroll through the parameters. Select
the parameter you would like to configure.
4.
When you have selected the desired parameter, change the parameter to the de-
sired value with the
Up button
and the
Down button
.
5.
In order to save the new parameter value, press and hold the
Function key
for
longer than 1 second.
Call up menu
Hold
Release
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
7.2.5
Configuring parameters of the adjustment menu
The following representation shows the available parameters and various configura-
tion options.
The
Adjustment
menu is open. See
Adjustment of the measuring input [
1.
Select the desired parameter you would like to configure.
2.
Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3.
The available configuration options are listed for each parameter in the following
representation.
Parameter
Values
Meaning
Zero point correction
T.OF
0.00
No zero point correction
-5.00 .. 5.00
Zero point correction in °C. and/or at °F -9.00 ..
9.00
Gradient correction of the temperature
T.SL
0.00
No gradient correction
-5.00 .. 5.00
Gradient correction in %