![GHM GREISINGER Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/ghm/greisinger/greisinger_operating-manual_2220427019.webp)
G 1501
Operation and maintenance | 6
B-H86.0.12.DB2-2.1
19 / 37
6.3.1.2
Buffer solutions
At least one buffer solution is required to calibrate the product. In the process, you
have the option of using a ready-to-use PHL buffer solution or mixing the solution
yourself with GPH buffer capsules - refer to the instructions.
Colour
10 °C
20 °C
25 °C
30 °C
40 °C
PHL 4.0
Red
4.02
4.00
4.01
4.01
4.01
PHL 7.0
Green
7.06
7.02
7.00
6.99
6.97
PHL 10.0
Blue
10.18
10.07
10.01
9.97
9.89
Ready-to-use buffer solutions in 250 ml dosing bottles with a dosing volume of 20 to 25 ml.
–
Plastic bottle
–
approx. 100 ml of distilled water
–
Buffer capsule
Colour
10 °C
20 °C
25 °C
30 °C
40 °C
GPH 4.0
Orange
3.99
3.99
4.01
4.01
4.03
GPH 7.0
Green
7.06
7.01
7.00
6.99
6.98
GPH 10.0 Blue
10.18
10.06
10.01
9.97
9.89
GPH 12.0 White
12.35
12.14
12.00
11.89
11.71
Buffer capsules for 100 ml buffer solution
1.
Fill a plastic bottle with approx. 100 ml of distilled water.
2.
Open the buffer capsule carefully by twisting the capsule halves and pulling. It
should be ensured that nothing is spilled. They can also be used without opening
them; opening the capsules only reduces to time for dissolving.
3.
Place the buffer capsule and its contents in the plastic bottle.
4.
Wait at least 3 hours.
5.
Shake well before using for the first time.
Then you can begin with calibration of the product.
Description
Prerequisite
Instruction
Outcome of an action