T45.0.0X.6C-09
Mounting and operating manual GMUD MP
Page 11 of 16
_____________________________________________________
_____________________________________________________________________________
4.3 Correction
Display
Values
Information
SET
Button ▼ and ▲
Offset correction of the measuring
Not active (factory
default)
…
Offset correction depending on display unit
Scale correction of the measuring
Not active (factory
default)
…
Scale correction in %
Peak detection of the measuring
Not active (factory
default)
Active:
Min.- and max.-value from non-averaged measuring value
If higher values are needed, please check sensor, if necessary return to manufacturer for
inspection.
Calculation: corrected value = (measured value
– OFFS) * (1+ SCAL /100)
4.4 Sea level correction (only for absolute devices)
Display
Values
Information
SET
Button ▼ and ▲
Sea level correction
Not active (factory
default)
Active
Altitude of sea level (only when
S.L ON
)
…
Altitude deviation to sea level in meters
Calculation formula for the sea level correction
Atmospheric conditions: international standard atmosphere
4.5 Transistor output (only available with option
–OUT)
Display
Values
Information
SET
Button ▼ and ▲
Lower alarm value (alarm when below value)
Not active (factory
default)
…
Output is active when measuring is below or equal this value
Always on
Upper alarm value (alarm when above value)
Not active (factory
default)
…
Output is active when measuring is above or equal this value
Always on