background image

E40.0.6x.6C-01 

Operating Manual EASY

LOG

 80BIN 

page 5 of 14 

 

3. Product description 

3.1  Required accessory 

The 

EASY

LOG 

80BIN

 is configurated, started and read-out via the 

EASYBus

-interface. 

Therefore the following accessories are required:  

 

Level converter

 

- RS232 <> 

EASYBus

 

(i.e. EBW 1, EBW 64, EBW 240) 

- USB <> 

EASYBus

 

(i.e. EBW 3) 

  Connecting cable: level converter to 

EASY

LOG

 

 

GSOFT 40K

 (version >7.14): 

 

  Windows software to start the logger and read out the logger-data 

 

EASYBus-Configurator

 (version >2.0): 

  for the configuration of the extended settings 

3.2  Notes on the logger’s delivery status 

The delivered logger is in a “sleep mode”, that means nothing is displayed and the 
power consumption is at its minimum. When there is a communication with the software 
for the first time, the EASY

LOG

 

leaves the sleep mode and ›Stop‹ is displayed. Then the 

device is ready for operation. 

Note: 

That sleep mode (nothing is displayed) cannot be restored again. 
“Stop“ or  “Halt“ are similar to the sleep mode. Both of  that operation 
modes ensure minimum consumption. 

Note: 

Logger start over keypad: 
before logger start over keypad, logger time must be adjust with 
GSOFT 40K. 

3.3  EASYBus connector 

 

 

 

 

 

 

 

 

 

 

 

 
 

Assignment of panel jack 

(view on pins) 

 
 
 
 

 

EASYBus

 connection

 

 
 
 
 
 
 
 
 
 
 
 

 

not connected

 

 

 

 

Содержание Greisinger EASYLOG 80BIN

Страница 1: ...RMANY Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual State data logger from version 2 4 EASYLOG 80BIN Please read these instructions carefully before use WEE...

Страница 2: ...es at the device logger operation 7 5 3 Operating possibilities via software GSOFT 40K 8 5 4 Operating possibilities via software EASYBus Configurator 8 6 Battery life time and recording period 8 7 Op...

Страница 3: ...and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the product 2 2 Safety signs...

Страница 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has t...

Страница 5: ...he extended settings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a commu...

Страница 6: ...ssibilities at the device display selection Function Procedure Display max value Push button shortly The max value is displayed for 4 seconds The max value display function is signalled by the display...

Страница 7: ...pressed for more than 2 min the menu is quit without saving Start recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with...

Страница 8: ...ing possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the configuration settings With that software the following adjustment are additionally p...

Страница 9: ...tatus messages are dis played STOP The EASYLOG recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The EASYLOG recording has been halt...

Страница 10: ...r to the manufacturer ERROR 7 The EASYLOG has detected a system error remedy Reset the error via GSOFT 40K If the error message is displayed furthermore please send the logger to the manufacturer for...

Страница 11: ...to 5 hours 18000 seconds Memory type cyclic logger recording can be activated by start stop function i e key press boundaries 64 independent recording sets are possible ring logger The logger records...

Страница 12: ...t and disposal 9 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or probe may be...

Страница 13: ...o not stop active logger has been activated at the configuration of the logger Remedy Deactivate this function by means of EASYBus configurator Problem The alarm settings cannot be changed Possible re...

Страница 14: ...ction diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Отзывы: