background image

E40.0.6x.6C-01 

Operating Manual EASY

LOG

 80BIN 

page 11 of 14 

 

8.  Specifications 

Input 

Passive volt-free switching contact 

  Measuring values 

1 = contact is closed (R < 50 Ohm) 
0 = contact is open (R > 20 kOhm) 

  Electrical connection 

approx. 0.5 m cable, with open ends 

Display 

two 4½-digit LCD-displays (about 7 and 5.8 mm high) 

Interface 

EASYBus 

  Busload 

equals to 2 EASYBus-elements 

Battery 

1x fixed lithium cell (Li / SOCl2) 3.6V / 1200 mAh (= 4.32 
Wh) 

  Battery life time 

depending on the chosen measuring cycle about 5 years 
(at 15 minutes and nominal temperature). 

Measured value memory 

250.000 measuring values 

Measuring cycle 

2 seconds to 5 hours (18000 seconds) 

Memory type: 

 

 

›cyclic logger‹ 

•  recording can be activated by start/stop function (i.e. key-

press, boundaries). 

•  64 independent recording sets are possible. 

 

›ring logger‹ 

The logger records continuously; 
as soon as the whole memory is used the old data of the 
beginning are overwritten. 

Recording period 

depending on the chosen measuring cycle 
i. e. 521 days at 3 min measuring cycle 
7.1 years at 15 min measuring cycle 

Alarm function 

alarm points selectable within the measuring range and 
alarm delay adjustable from 0 to 500 min 

Nominal temperature  

25 °C 

Operating temperature 

-

25.0 … +60.0 °C 

Storage temperature 

-

30.0 … +70.0 °C 

Housing 

ABS housing, transparent screen made of polyester 

  Dimensions 

48.5 × 48.5 × 35.5 mm  (L × W × D) 
without sensor and plug 

  Protection rate 

IP65 

(IP00 for open cable ends of input cable)

 

 

 

 

Содержание Greisinger EASYLOG 80BIN

Страница 1: ...RMANY Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual State data logger from version 2 4 EASYLOG 80BIN Please read these instructions carefully before use WEE...

Страница 2: ...es at the device logger operation 7 5 3 Operating possibilities via software GSOFT 40K 8 5 4 Operating possibilities via software EASYBus Configurator 8 6 Battery life time and recording period 8 7 Op...

Страница 3: ...and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the product 2 2 Safety signs...

Страница 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has t...

Страница 5: ...he extended settings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a commu...

Страница 6: ...ssibilities at the device display selection Function Procedure Display max value Push button shortly The max value is displayed for 4 seconds The max value display function is signalled by the display...

Страница 7: ...pressed for more than 2 min the menu is quit without saving Start recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with...

Страница 8: ...ing possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the configuration settings With that software the following adjustment are additionally p...

Страница 9: ...tatus messages are dis played STOP The EASYLOG recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The EASYLOG recording has been halt...

Страница 10: ...r to the manufacturer ERROR 7 The EASYLOG has detected a system error remedy Reset the error via GSOFT 40K If the error message is displayed furthermore please send the logger to the manufacturer for...

Страница 11: ...to 5 hours 18000 seconds Memory type cyclic logger recording can be activated by start stop function i e key press boundaries 64 independent recording sets are possible ring logger The logger records...

Страница 12: ...t and disposal 9 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or probe may be...

Страница 13: ...o not stop active logger has been activated at the configuration of the logger Remedy Deactivate this function by means of EASYBus configurator Problem The alarm settings cannot be changed Possible re...

Страница 14: ...ction diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Отзывы: