8. Specifications
Input
Passive volt-free switching contact
Measuring values
1 = contact is closed (R < 50 Ohm)
0 = contact is open (R > 20 kOhm)
Electrical connection
approx. 0.5 m cable, with open ends
Display
two 4½-digit LCD-displays (about 7 and 5.8 mm high)
Interface
EASYBus
Busload
equals to 2 EASYBus-elements
Battery
1x fixed lithium cell (Li / SOCl2) 3.6V / 1200 mAh (= 4.32
Wh)
Battery life time
depending on the chosen measuring cycle about 5 years
(at 15 minutes and nominal temperature).
Measured value memory
250.000 measuring values
Measuring cycle
2 seconds to 5 hours (18000 seconds)
Memory type:
›cyclic logger‹
• recording can be activated by start/stop function (i.e. key-
press, boundaries).
• 64 independent recording sets are possible.
›ring logger‹
The logger records continuously;
as soon as the whole memory is used the old data of the
beginning are overwritten.
Recording period
depending on the chosen measuring cycle
i. e. 521 days at 3 min measuring cycle
7.1 years at 15 min measuring cycle
Alarm function
alarm points selectable within the measuring range and
alarm delay adjustable from 0 to 500 min
Nominal temperature
25 °C
Operating temperature
-
25.0 … +60.0 °C
Storage temperature
-
30.0 … +70.0 °C
Housing
ABS housing, transparent screen made of polyester
Dimensions
48.5 × 48.5 × 35.5 mm (L × W × D)
without sensor and plug
Protection rate
IP65
(IP00 for open cable ends of input cable)