Before installing the radiometer, refill the cartridge containing silica-gel crystals. Silica 
gel absorbs humidity in the dome chamber and prevents, in particular climatic condi-
tions, condensation on the internal walls of the domes and measurement alteration.  

Do not touch the silica gel crystals with your hands while refilling the cartridge. Carry 
out the following instructions in an environment as drier as possible: 

1.

 

Loosen the three screws that fix the white shade disk. 

2.

 

Unscrew the silica gel cartridge using a coin. 

3.

 

Remove the cartridge perforated cap. 

4.

 

Open the sachet containing silica gel (supplied with the radiometer). 

5.

 

Fill the cartridge with the silica gel crystals. 

6.

 

Close  the  cartridge  with  its  own  cap,  paying  attention  that  the  sealing  O-ring  be 
properly positioned.  

7.

 

Screw the cartridge to the radiometer body using a coin. 

8.

 

Check that the cartridge is screwed tightly (if not, silica gel life will be reduced). 

9.

 

Position the shade disk and screw it with the screws. 

10.

 

The radiometer is ready for use. 

The figure below shows the operations necessary to fill the cartridge with the silica gel 
crystals. 

 

Fig. 3.1: filling the silica-gel cartridge 

Silica-gel 

cartridge 

Perforated cap 

Sealed sachet 
of silica-gel crystals 

Filling 

Closing the cartridge 

Содержание DeltaOHM LPUVA02

Страница 1: ...Operating manual UVA Radiometer LPUVA02 www deltaohm com English Keep for future reference...

Страница 2: ...ON 5 4 ELECTRICAL CONNECTIONS 8 4 1 LPUVA02 CONNECTIONS 8 4 2 LPUVA02AC CONNECTIONS 9 4 3 LPUVA02AV CONNECTIONS 9 5 MEASUREMENT 10 5 1 LPUVA02 10 5 2 LPUVA02AC 10 5 3 LPUVA02AV 10 6 MAINTENANCE 11 7 T...

Страница 3: ...VOLTAGE output to be de fined when ordering 0 200 W m2 Each radiometer is individually calibrated at the factory and is distinguished by its own calibration factor The calibration is performed by com...

Страница 4: ...ined thanks to the par ticular shape of the diffuser and of the housing The deviation between the theoretical response and the measured one is shown in the Fig 2 2 Angle Fig 2 2 The excellent relation...

Страница 5: ...a coin 3 Remove the cartridge perforated cap 4 Open the sachet containing silica gel supplied with the radiometer 5 Fill the cartridge with the silica gel crystals 6 Close the cartridge with its own c...

Страница 6: ...lf For fixing use the holes on the radiometer body remove the shade disk to access the holes and reposition it after mounting or the suitable accessories see the fig ures below In order to allow an ac...

Страница 7: ...LPUVA02 7 V2 0 HD2003 79K HD2003 77C HD2003 83 HD2003 83 1 HD2003 85K Fig 3 3 fixing accessories...

Страница 8: ...to the connector 4 1 LPUVA02 CONNECTIONS The radiometer LPUVA02 is passive and does not require power supply It is to be connected either to a millivoltmeter or to a data acquisition system Typically...

Страница 9: ...rdered output and requires external power supply 10 30 Vdc for 0 1 V and 0 5 V outputs 15 30 Vdc for 0 10 V output It is to be connected to a power sup ply and an instrument with voltage input as show...

Страница 10: ...PUVA02AC The 4 20 mA output signal corresponds to 0 200 W m2 irradiance range The irradiance Ee is obtained by measuring with a multimeter the current Iout absorbed by the sensor and applying the foll...

Страница 11: ...is case the performed reading is highly over estimated To minimize the condensation the radiometer is provided with a car tridge containing dessicant material silica gel The efficiency of the silica g...

Страница 12: ...93 nm 1 10 320 nm 400 nm 1 100 Peak 365 nm Response time 0 5 s 95 Output LPUVA02 V Wm 2 LPUVA02AC 4 20 mA LPUVA02AV 0 1 0 5 0 10 V depending on the model Power supply 10 30 Vdc 15 30 Vdc only for 0 10...

Страница 13: ...safety can be ensured only if all standard safety measures as well as the specific measures described in this manual are followed Do not use the instruments in places where there are Corrosive or flam...

Страница 14: ...the radiometer on 40 mm clamping To install the radiometer on a transverse mast LPS6 Kit for the installation of the radiometer The kit includes 750 mm mast base fitting graduated support plate brack...

Страница 15: ......

Страница 16: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Отзывы: