LPPYRA03
- 12 -
V2.8
4.6
LPPYRA03ACS
CONNECTIONS
The pyranometer LPPYRA03ACS has two outputs:
•
One
4…20 mA
output, requiring
10…30 Vdc
external power supply. It is to be
connected to a power supply and an instrument with 4…20 mA input as shown in
fig. 4.6. The load resistance of the instrument reading the signal must be
≤ 500 Ω
.
•
One
RS485 Modbus-RTU
output, requiring
5…30 Vdc
external power supply.
It is to be connected to a power supply and to a PLC, a data logger or a
RS485/USB or RS485/RS232 converter for PC as shown in fig. 4.6. The RS485
output is not isolated.
Connector
Function
Color
1
Power supply negative (GND)
Blue
2
Power supply positive (+Vdc)
Red
3
Digital and analog ground
(SGND)
Black
4
RS485 A/-
Brown
5
RS485 B/+
White
6
Housing / Cable shield (SH)
Black (thick wire)
7
Analog output positive (AOUT)
Green
8
Not connected
R
220
T
Ω
Shield
Shield
Lmax = 1200m
B/+
A/-
1
2
3
4
5
6
7
8
B/+
A/-
SGND
+Vdc
GND
4...20 mA
AOUT
SGND
GND
+Vdc
A/-
B/+
SH
R
220
T
Ω
Fig. 4.6: LPPYRA03ACS[4] connections
Before connecting the pyranometer to the RS485 network, set the address and the
communication parameters, if different from the factory preset (see chapter 6).
Power supply
CPM12-8DA… cable
Pyranometer M12
male connector
PLC, data logger or
RS485/USB or
RS485/RS232
converter for PC
Other sensors with
RS485 output
Connect to ground only if it is not possible to
ground locally the housing of the pyranometer
Содержание Delta OHM LPPYRA03
Страница 1: ...Operating manual Pyranometer LPPYRA03 www deltaohm com English Keep for future reference...
Страница 7: ...LPPYRA03 7 V2 8 Fig 3 3 fixing accessories...
Страница 25: ...LPPYRA03 25 V2 8 NOTES...
Страница 26: ...LPPYRA03 26 V2 8 NOTES...
Страница 27: ......