![GHM Delta OHM HD2304.0 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/ghm/delta-ohm-hd2304-0/delta-ohm-hd2304-0_operating-manual_2220403014.webp)
HD2304
-
14
-
V1.8
D
I RECT CON N ECTI ON OF
4 -
W I RE
Pt 1 0 0
SEN SOR
4
w ir e Pt 1 0 0 se n sor s
can be
soldered direct ly t o t he pins of
t he fly ing fem ale connect or
wit hout m ak ing use of t he TP47
board. The 4 w ires of t he Pt 100
sensors have t o be soldered as
indicat ed in t he figure on t he left .
I n order t o use t his t ype of probe
it is necessary t o set up t he
m enu it em “ Probe Type” as
described at page 10.
The P100 probe is recognized
upon t urning on t he inst rum ent :
connect t he probe when t he
inst rum ent is sw it ched off and
t hen t urn it on.
Se n sor
D ir e ct solde r in g t o t h e
conn e ct or
Pt 100 4
wires
View of t he solder ing side
of t he fly ing fem ale connect or