Error messages
Display
Description
Broken
line
Pt100
Please check
Measured val.
Pt100
Please check
> 650
°
C
Measured val.
Pt100
Please check
< -120
°
C
Device
temperature
Internal error
Please check
Memory
error
Fatal error
Replace device
Hardware
error
Fatal error
Replace device
Relay
fault
Fatal error
Replace device
A cable interruption was detect at the terminals of the Pt100.
Check connection lines and Pt100 probes.
The measurement exceeds the maximum temperature for Pt100
probes.
The measurement undercuts the minimum temperature for Pt100
probes.
The internal temperature of the device has detected an impermissi-
bly high or low value. Check the installation conditions and envi-
ronmental temperature of the device.
Monitoring of the program- and/or working memory revealed an er-
ror.
The device must be inspected in the factory.
A deviation was detected for an internal voltage.
The device must be inspected in the factory.
An incorrect setting of the relay contact was detected.
The device must be inspected in the factory.