background image

4

463X Chlorine Systems Manual

Follow the steps below to set up a new Chlorine Analyzer System. Refer to the individual component manuals for detailed information.

Step 1. 

Mount the panel on a vertical 

fl

 at surface using appropriate hardware.

Step 2. 

Open the wiring enclosure and wire input power 

(see page 8: Wiring Input, page 9: Wiring Output, and page 10: Electrcial Box Wiring Schematic).

Step 3. 

Wire any 4 to 20 mA and relay output.

Step 4. 

Remove sensor access plugs from the 

fl

 ow cell (Figure 1). 

If the optional pH sensor is NOT used, do not remove the left-side plug from the 

fl

 ow cell.

Step 5. 

Remove the protective cap from the chlorine electrode. 

(Keep the electrode cap in a safe place for future use. It is recommended to use the cap to protect the sensor during 

the removal of the electrode for cleaning or maintenance of the 

fl

 ow cell).

Step 6.  Complete Sensor Preparation

 (see page 11) and install the chlorine sensor into the electronics 

(see page 12: Sensor Installation). Install the chlorine electrode into the 

fl

 ow cell. 

The chlorine sensor is installed in the right-side access port, optional pH sensor is installed in the left-side access port.

NOTE: 

All new chlorine and pH sensors require calibration during the start up of a system and throughout the life of the sensor. 

A new chlorine sensor requires a conditioning period of up to 4 hours with power on and chlorinated water 

fl

 owing past 

the sensor prior to calibration. See page 24: Chlorine Sensor Calibration, for chlorine calibration and set up procedure.

NOTE:

  If a Free Chlorine sensor without optional pH sensor is used, pH value must be 

"hard-coded" into the system (See page 23: Manual pH compensation). 

If optional pH sensor is installed, see page 22 to complete pH Sensor Calibration.

Step 7. 

Repeat step 5 and 6 if the optional pH sensor is being used.

Step 8. 

Install the in

fl

 uent water source to the "Inlet Port" nipple assembly of the 

fl

 ow 

cell. Install 3/8-inch tubing and secure with a hose clamp. 

(Not included. See page 13: Tubing Connections)

Step 9. 

Install 3/8-inch tubing and secure with a hose clamp on the "Drain" port and 

direct the tube to a proper drain. (Not included)

Step 10. 

Verify the inlet and drain ball valves are in the open position and 

the sample port is in the off position (See page 13: Tubing Connections).

Step 11. 

Turn on the in

fl

 uent water source and check the system for leaks.

Step 12. 

Apply power to the system and allow system to initialize. Calibrate per 

instructions 

(See page 24: Chlorine Sensor Calibration and page 22: pH Sensor Calibration).

NOTE:

  As factory default, the 4630-X Chlorine Panel Assembly is set to measure Free Chlorine. If a 2632 Chlorine Dioxide 

sensor is to be used to convert to a Chlorine Dioxide Panel Assembly, refer to the 8630-3 OPTIONS Menu on page 20.

Do not turn on power at this time.

1.  Chlorine panel assembly

 

a.  Built-in pressure regulator 15 to 120 psi.

 

b.  3/8 inch hose barb connectors.

2.  1 each 2630 series Free Chlorine sensor or 2632 Chlorine Dioxide sensor; each with protective cap

 

a.  1 spare membrane cap

 

b.  2 bottles of electrolyte solution

 

c.  1 syringe needle (taped to bottle)

 

d.  1 syringe

3.  1 each Flat pH sensor 3-2724-00 (159 001 545) (3-4630-11, -21, -31 and 3-4632-11 models only)

4.  Manual package

 

a.  English panel assembly manual

 

b.  CD with multi-language manuals

 

c.  Wall mounting hardware

 

d.  Drill template

5.  1 each North American Type B power cord

6.  Customer supplied

 

a.  3/8" hose input and drain

 

b.  Hose clamps

Sensor access plugs 

Figure 1

  Quick Start

  System Inventory

Содержание Signet 4630

Страница 1: ...The 4630 Free Chlorine and 4632 Chlorine Dioxide systems can be used for reporting chlorine residuals in accordance with EPA Method 334 0 Complete chlorine analyzer system allows quick setup and easy...

Страница 2: ...onal Protective Equipment PPE Always utilize the most appropriate PPE during installation and service of Signet products Pressurized System Warning Sensor may be under pressure take caution to vent sy...

Страница 3: ...Rating 1 to 8 bar 15 to 120 psi Pressure Regulator 0 69 bar 10 psi variation over all ranges of flow and pressure Flow tolerance 15 or rated specification above Flow rate limits 8 to 12 gph US 30 24 t...

Страница 4: ...Manual pH compensation If optional pH sensor is installed see page 22 to complete pH Sensor Calibration Step 7 Repeat step 5 and 6 if the optional pH sensor is being used Step 8 Install the influent...

Страница 5: ...low rate indicator Flow range limits Flow regulator with strainer Inlet port Drain port 3 8 inch hose Sampling port pH sensor electronics 3 2750 7 159 001 671 pH electrode 3 2724 00 159 001 545 Quick...

Страница 6: ...Flat washer Disc spring curved washer Knurled thumb nut 1700 2512 Flow cell bolt 2490 0554 Included in kit 3 4630 390 Item Quantity Flow cell nut 6 Flat washer 15 Disc spring 9 Knurled thumb nut 3 Flo...

Страница 7: ...the panel according to local electrical building and or plumbing codes and seismic requirements Use four 6 mm in diameter screws or bolts of sufficient length to mount the panel to a sturdy vertical...

Страница 8: ...Wiring Open the electrical box by loosening the four corner screws of the front cover Insert input power wiring into the pre drilled access hole on the left side of the electrical box using appropriat...

Страница 9: ...transmitter Recommended torque for the terminals is 7 lb in Do not run 4 to 20 mA loop cables in the same conduit as the power or other high voltage wiring Remove one installed jumper wire from both o...

Страница 10: ...net Chlorine Sensor Electronics 3 2650 7 159 001 670 10 9 8 7 6 5 4 3 2 1 GND 10 9 8 7 6 5 4 3 2 1 GND NO COM RELAY 2 NC NO COM RELAY 1 NC LOOP 2 LOOP 2 LOOP 1 LOOP 1 EARTH GROUND Black White Yellow R...

Страница 11: ...quire different electrolyte solutions 2630 Free Chlorine Electrode 2632 Chlorine Dioxide Electrode NOTE Inspect and change the membrane cap if damaged See page 36 Maintenance for cleaning procedure A...

Страница 12: ...e flow cell DryLoc Connection to Sensor Electronics Locking Ring Locking Ring Lock Lock 3 2724 00 159 001 545 pH electrode 3 2630 X FCL electrode 3 2750 7 159 001 671 pH electronics 3 2650 7 159 001 6...

Страница 13: ...sitioned lower than the influent water source to allow proper flow through the flow cell The flow cell must drain by gravity not system pressure When testing allow the sample to flow for a few seconds...

Страница 14: ...the previous operating display When editing the CALIBRATE or OPTIONS menus the transmitter will return to the default display after 10 minutes and then the VIEW menu in another 10 minutes if no activi...

Страница 15: ...ering the Key Code the display will show the first item in the selected menu Step 3 Scroll the menu in a loop with the or arrow keys Step 4 Press the key to select the menu item to be edited The first...

Страница 16: ...active during editing Only the flashing element can be edited Press ENTER to save the new value When you have set your desired value pressing the ENTER key stores the value on the screen making it imm...

Страница 17: ...maximum values for the current loop output Units are derived from Loop 1 Source If the Source 1 is changed be sure to re edit this page Loop 2 Source pH Choose pH or chlorine for this 4 to 20 mA curre...

Страница 18: ...Range Relay1 Rng ppm 0 00 5 00 Enter the range where Relay 1 will activate above and below this setpoint Relay1 PlsRate 120 pulses min Set the maximum pulse rate Range 1 to 400 pulses min Relay 1 Mod...

Страница 19: ...o Last Cal 10 18 09 Edit the calibration date End of Calibrate Menu Relay 2 Rng ppm 0 00 5 00 Enter the range where Relay 2 will activate above and below this setpoint Relay 2 Hys 0 20 ppm This relay...

Страница 20: ...OWN keys to manually output any current value from 3 6 mA to 21 00 mA to test Loop 1 output Test Loop 2 Press UP or DOWN keys to manually output any current value from 3 6 mA to 21 00 mA to test Loop...

Страница 21: ...of operation Low High C xxxx x xxxx x Lowest and highest temperatures the Cl sensor has been subjected to during operation Test Relay 1 Press UP or DOWN keys to manually toggle Relay 1 Off and On The...

Страница 22: ...5 prior to initially installing the sensor and during its normal lifetime If a pH sensor is not available but pH determination is necessary measure process pH with a separate test and enter the value...

Страница 23: ...nse it in distilled water 2 Place the pH sensor in a different buffer solution example pH 4 01 The pH standard and slope must be at least 2 pH units apart 3 Note the pH and mV readings on the View Men...

Страница 24: ...int E X A M P L E ENTER ENTER Chlorine Sensor Calibration Zero Point Calibration Zero Point Calibration is only used at the factory to program the sensors zero calibration point This value does not sh...

Страница 25: ...nd relay functions if applicable NOTE The current and relay outputs can be tested in the Options Menu Current Loop Settings Current outputs are passive outputs that can be spanned in the forward and r...

Страница 26: ...l The output will be 100 pulses minute when the input value is 1 or less The starting point endpoint and maximum pulse rate are selectable in the CALIBRATE menu Mechanical Relay Functions The 8630 rel...

Страница 27: ...11 Scroll down to the Relay 1 Hys menu 12 Press to enter Edit Mode 13 Set the hysteresis dead zone for this relay This affects the turn off only 0 3 ppm 14 Press Enter 15 Scroll down to the Relay 1 D...

Страница 28: ...lorine Gnd I O V 18 17 16 Signal Ground Digital S3 L data 5 VDC White Red Black Shield pH Signet 3 2750 7 159 001 671 N C 8630 Chlorine Gnd I O V 15 14 13 Signal Ground Digital S3 L data 5 VDC White R...

Страница 29: ...moval a Disconnect flow through system b Drain below sensor level 3 Confirm chemical compatibility before use 4 Do not alter product construction These electrodes require the Signet 2650 Amperometric...

Страница 30: ...e is 7 0 and the pH variation is 0 2 then automatic pH compensation is not required pH variation 6 5 7 0 7 5 8 0 8 5 9 0 0 0 3 0 1 0 2 Sample pH Automatic pH compensation recommended in ranges within...

Страница 31: ...ng with chemicals or solvents 4 Place a Lockout tag on the pipe when the sensor is removed for maintenance to prevent accidental opening and exposure to potentially hazardous chemicals Lubricate O rin...

Страница 32: ...buffer solutions from the electrode Dispose of all buffer solutions at the end of the day If the pH sensor will not calibrate within acceptable limits clean the electrode and recalibrate If the calibr...

Страница 33: ...for the application be used during the reassembling of the ow cell Maintenance 463X Chlorine Analyzer Flow Cell CAUTION Over time a sensor can get tight in the flow cell fitting When removing the sens...

Страница 34: ...for special O ring fitting instructions Butt all O ring joints together so there is no gap Remove the O rings during flow cell disassembly Both the O rings and all sealing grooves should be examined f...

Страница 35: ...shers and nuts 7 Install the assembled pressure regulator and the two round O rings into the flow cell 1 Install the four O ring segments 2 Insert the center bolt into the block 3 4 Align and place th...

Страница 36: ...olution Slowly fill until solution begins to flow out of holes Do not allow the solution to run down the electrode and wet the electrical contacts in the DryLoc connector 6 Slowly screw on the membran...

Страница 37: ...e nA reading should start to rise Response time and nA reading will depend upon the temperature of the appropriate solution 15 Once the sensor s nA reading reaches approximately 300 360 nA allow the s...

Страница 38: ...the glass wet when out of the process Never store the electrode at temperatures below 0 C 32 F or allow it to dehydrate Never scrape or sand the glass electrode surface Treat glass electrode surfaces...

Страница 39: ...tmosphere Algae growth in flow cell Low chlorine concentration in water Locate flow cell away from light Exposure to light Excessive small bubbles in flow cell accumulating on sensors Water source con...

Страница 40: ...water Electrode not properly conditioned Noise interfering with the measurement Calibration incorrect Condition new sensor for 4 hours Cap replacement or electrolyte refill 2 hours Properly ground sys...

Страница 41: ...e calibration point is too close to standard point Wrong data is entered pH buffer solution used has value too close to standard point Re enter correct data Use proper buffer solution at least 2 pH un...

Страница 42: ...ged pH sensor Check pH sensor connections or change pH input to Manual in Options Menu FCl CHK pH PREAMP pH preamp is not detected pH preamp is not connected Wrong wiring Damaged preamp Check pH pream...

Страница 43: ...priate DPD test Electrode not making good contact with electronics Inspect and reconnect Unstable output from sensor Electrode needs to be repolished Polish gold electrode Not enough electrolyte Check...

Страница 44: ...on of Conformity according to FCC Part 15 This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2...

Страница 45: ...y not cause harmful interference and 2 This device must accept any interference received including interference that may cause undesired operation General Compatibility Electrode Signet 2724 00 Flat p...

Страница 46: ...02 to 2 ppm mg L 3 2630 2 0 05 to 5 ppm mg L 3 2630 3 0 1 to 20 ppm mg L Free Chlorine pH 5 0 to 8 5 pH Chlorine Dioxide 3 2632 1 0 to 2 ppm mg L Chlorine Dioxide pH 4 0 to 11 0 pH Temperature 0 C to...

Страница 47: ...149 F 0 to 6 9 bar 0 to 100 psi 65 C to 85 C 149 F to 185 F Linearity derated 6 9 to 4 0 bar 100 to 58 psi Temperature Storage 0 C to 50 C 32 F to 120 F CAUTION The electrode glass will be shattered i...

Страница 48: ...ectrode pH flat PT1000 in NPT 3 2750 7 159 001 671 pH Inline Electronics Digital S3 L 4 6 m 15 ft cable 3 2650 7 159 001 670 Chlorine Inline Amperometric Electronics Digital S3 L 4 6 m 15 ft cable Acc...

Отзывы: