GF 263X Series Chlorine Electrodes 

IN

OUT

pH

Cl

IN

OUT

pH

Cl

Installation

Lubricate O-rings with a  

non-petroleum based, 

viscous lubricant (grease) 

compatible with the system. 

Align tab on electrode with 

notch on fitting. Insert until 

fully seated.

Do Not Use Lubricant or Sealing Tape on Threads.

Do Not Overtighten. Do Not Use Tools.

pipe tee

pipe tee

pipe tee

pipe tee

When using the chlorine 

panel flow cell it is not 

necessary to lubricate 

this O-ring.

Mounting Position - PVC Tee

Best

Flow Direction

Good

45º

or

or

45º

Mount the flow cell where the sensor will be easily accessible.

To avoid air bubble entrapment, do not mount with downward flow.

NO

Flow

Содержание GF 3-2630-X

Страница 1: ...ly and a secure connection Gold plated contacts and an O ring seal ensure a waterproof and reliable interconnect to the 2650 Amperometric Electronics Integrated temperature element for automatic tempe...

Страница 2: ...ts user to risk of potential damage to product by ESD Personal Protective Equipment PPE Always utilize the most appropriate PPE during installation and service of GF products Pressurized System Warnin...

Страница 3: ...ignet com for details Declaration of Conformity according to FCC Part 15 This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not c...

Страница 4: ...sink is recommended 1 Remove the protective bottle from the end of the electrode 2 Remove the membrane cap from the front of the sensor Note When new sensors are shipped the membrane cap is not tight...

Страница 5: ...le operating pressure must be less than 0 48 bar 7 psi Higher pressures will damage the electrode The electrode should not be used in water containing surfactants oils organic chlorine or stabilizers...

Страница 6: ...injecting with electrolyte solution Slowly injecting the electrolyte solution into the sensor to avoid introducing air bubbles The electrode holds approximately 14 milliliters of solution Slowly fill...

Страница 7: ...bottle with tap water added Storage periods more than 1 week Remove the membrane cap and internal electrolyte solution Rinse the sensor internal chamber with DI water or cold tap water drain and allow...

Страница 8: ...7 11 Fill beaker with a 12 mm inch of the appropriate solution 12 Position or suspend the sensor 6 mm to 12 mm in to in above the appropriate solution DO NOT SUBMERGE THE SENSOR See Figure 2 13 Apply...

Страница 9: ...t until fully seated Do Not Use Lubricant or Sealing Tape on Threads Do Not Overtighten Do Not Use Tools pipe tee pipe tee pipe tee pipe tee When using the chlorine panel flow cell it is not necessary...

Страница 10: ...tions which help determine the specific application The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 h...

Страница 11: ...resent in water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outsi...

Страница 12: ...1 to 20 ppm mg L 3 2632 1 159 001 767 Chlorine Dioxide electrode 0 02 to 2 ppm mg L Accessories and Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 m...