background image

6

263X Series Chlorine Electrodes

 10. Troubleshooting

Problem

Possible Causes

Remedies

Sensor cannot be calibrated.
Output is higher than DPD test
(out of range).

Sensor conditioning time too short

Condition for 4 hours minimum prior to initial 
calibration

Interference from contaminants

See technical data

Membrane cap damaged

Replace cap and recondition

DPD chemicals bad

Use fresh reagents

pH outside of working range

See technical data

Sensor cannot be calibrated.
Output is lower than DPD test.

Sensor conditioning time too short

Condition for 4 hours minimum prior to initial 
calibration

Chlorine content too low

DPD value must be greater than 0.5 ppm to 
calibrate

Low 

fl

 ow rate

Check to make sure 

fl

 ow rate is suf

fi

 cient

Air bubbles on electrode membrane

Inspect visually. Tap to remove bubbles. Mount 
at an angle.

Low or no electrolyte in electrode

Fill electrode with electrolyte

Organic chlorination agents present in water

See technical data

Surfactants in water

Remove surfactants and replace cap

Membrane cap coated

Clean or replace membrane cap

Membrane cap loose

Tighten or replace membrane cap

pH outside working range.

See technical data

Sensor output very low

Sensor conditioning time too short

Condition for 4 hours minimum prior to initial 
calibration

Chlorine content too low

Add chlorine to validate

Only bound chlorine present. No free chlorine

Check for chloramine with appropriate DPD test

Electrode not making good contact with electronics Inspect and reconnect

Unstable output from sensor

Air bubbles on electrode membrane

Inspect visually. Tap to remove bubbles. Mount 
at an angle.

Membrane damaged

Replace membrane. Condition sensor for at 
least 2 hours and recalibrate.

Electrode not making good contact with electronics Inspect and reconnect

Non-sensor problem

Check 3-2650 Electronics connection to 
electrode (see 3-2650 manual for instructions). 
Make sure connections are dry. Check 
instrument hookup.

Transmitter error messages related to calibration are detailed in the Signet 8630 Chlorine Transmitter operation manual.

Содержание Signet 3-2630 Series

Страница 1: ...vice time chlorine range high and low pH with optional Signet pH electrode temperature limits and more are stored on the chip This information is accessible via the Signet 8630 transmitter Signet s pa...

Страница 2: ...rane caps carry no warranty Diluted HCl can irritate the eyes and skin use proper safety equipment Do not use surface tension reducing chemicals detergents or solvents on the membrane Electrode Range...

Страница 3: ...greater than 2 months 4 If the sensor still fails to work properly continue with steps 11 17 Storage periods greater than 2 months If the sensor has been in storage for a long period of time or used...

Страница 4: ...MERGE THE SENSOR See Figure 2 13 Apply power to the system 14 Monitor the nA of the sensor press the down arrow once on the 8630 transmitter The nA reading should start to rise Response time and nA re...

Страница 5: ...The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 hypochlorous acid HOCl and hypochlorite OCl Above pH...

Страница 6: ...n water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outside worki...

Страница 7: ...Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 mL 3 2632 391 159 310 160 Chlorine Dioxide electrolyte solution 30 mL 3 2630 394 159 310 164 Free Chl...

Страница 8: ...trode range Sensor Conditioning New first start up 4 hours maximum before calibration Subsequent start ups 2 hours maximum Temperature Element PT1000 Operational Ranges and Limits Free Chlorine Range...

Отзывы: