background image

8

2850 Conductivity/Resistivity Sensor Electronics

EasyCal is a single-point calibration system.

 During this procedure, if the measured value is within ±10% of any of the test 

values listed below, the 2850 will automatically recognize the test value and calibrate the output to that value.

NOTE:  

The 

fi

 rst step (Reset) is recommended each time an electrode is replaced, but is NOT necessary upon initial installation or 

periodic calibration.

1. Reset 

the 2850 Sensor to factory calibration: Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Apply power, wait 

at least 3 seconds, then press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds). 
When the LED goes off, release SW1; reset is complete. Reset SW3 and SW4 to the proper settings.

2. Place

 the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the 

electrode to dislodge any air bubbles visible on the surface of the electrode.

• 146.93 

μ

S, 1408.8 

μ

S, 12856 

μ

S (per ASTM D1125-95)

• 10 

μ

S, 100 

μ

S, 200 

μ

S, 500 

μ

S, 1000 

μ

S, 5000 

μ

S, 10,000 

μ

S, 50,000 

μ

S, 100,000 

μ

S.

Allow at least 2 minutes for the electrode response to stabilize.

3. Press

 and hold SW1 for approximately 8 to 10 seconds. During this time, the LED will come on steady then go back off. 

(If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section).

Calibration is complete. Return the system to service.

EasyCal

Dual Input Calibration

Dual Input 3-2850-63 Electronics

•  The 2850-63 can only be calibrated if both inputs are used, or if Input 2 is disabled.

•  To disable Input 2: Set SW4, #4 to OPEN.

•  Place each sensor into a test solution that represents the operating range of that sensor. When SW1 is pressed, the 2850 will 

calibrate both inputs simultaneously.

•  Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant. 

Place the larger cell constant on Input 1

. This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN).

Calibration

•  All 2850 Conductivity/Resistivity Sensor Electronics are factory calibrated for maximum out-of-the-box accuracy.
•  The EasyCal feature allows calibration to be performed at the sensor.

a.  SW2: 4 to 20 mA output range selection switch bank
b.  Power and Output terminal block: 4 to 20 mA or 

Digital (S

3

L) Outputs

c.  SW3 and SW4: Electrode Cell Constant selection 

switch banks

d.  D3: LED indicator
e.  SW1: EasyCal Push-button
f. 

Input 1 & Input 2: Conductivity Electrode connections

SW1

SW2

SW3

SW4

D3 

ON 

CTS 

ON 

CTS 

ON 

CTS 

Input 1

Input 2

e

d

b

a

c

f

The 2839-2842 Conductivity sensors come with a certi

fi

 cate 

of calibration. This certi

fi

 cate provides information regarding 

the actual cell constant and temperature  offset as tested and 
veri

fi

 ed according to NIST standards. This information has 

been programmed into the 2850 electronics as its default 
factory value. If required, the sensor can be recerti

fi

 ed by 

contacting the GF Signet repair department.

NOTE:

When the electrode is changed from its original factory 
provided electrode or the application requires a cell constant 
change, reset SW3 or SW4 switches (see page 6). 
Then perform a wet calibration with calibration standards.

Содержание Signet 2850

Страница 1: ...vity test solution values for simple field calibration The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the fact...

Страница 2: ...t accept any interference received including interference that may cause undesired operation Specifications Caution Warning Danger Indicates a potential hazard Failure to follow all warnings may lead...

Страница 3: ...li Cleaners Acids Bases 0 01 Cell 0 10 Cell 1 0 Cell 10 0 Cell 20 0 Cell 3 2841 3 2842 26 8 mm 1 13 in 41 9 mm 1 65 in 3 2823 203 mm 8 0 in 89 mm 3 5 in 3 2839 59 3 mm 2 33 in 73 7 mm 2 90 in 3 2840 2...

Страница 4: ...sanitary flange fittings must be installed using this option While the 2850 5X and 2850 6X electronics cannot be submerged either model will accommodate tank installation Select any electrode with a...

Страница 5: ...q Input 2 Red S L Red GND White Shield 5VDC Black S L Red GND White Shield 3 3 Analog Output 1 Analog Output 2 if applicable if applicable Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input...

Страница 6: ...W3 or SW4 Integral 2850 Systems Single Input The 2850 51 XX and 2850 52 XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single input electronic...

Страница 7: ...to 500 S 0 to 1000 S O C C C 10 to 2 M 5 to 0 S 50 to 0 S 500 to 0 S 1 000 to 0 S C O C O 0 to 2 M 0 to 10 S 0 to 100 S 0 to 1 000 S 0 to 2 000 S C O C C 2 to 0 M 10 to 0 S 100 to 0 S 1 000 to 0 S 2...

Страница 8: ...ed if both inputs are used or if Input 2 is disabled To disable Input 2 Set SW4 4 to OPEN Place each sensor into a test solution that represents the operating range of that sensor When SW1 is pressed...

Страница 9: ...e test solution is outside the 10 tolerance The electronics or the electrode is defective Use fresh test solutions and restart the calibration Replace the electronics or the electrode After completing...

Страница 10: ...1 Turn the power off 2 Set SW3 to accept a 0 01 cell constant Leave SW4 in the all OPEN position 3 Install the simulator into the 2850 at the Input 1 conductivity electrode terminals 4 Connect the 285...

Страница 11: ...11 2850 Conductivity Resistivity Sensor Electronics Notes...

Страница 12: ...tem Digital S3 L output 0 01 cell ISO threads 3 2850 51 40D 159 001 344 Integral 2850 system Digital S3 L output 0 1 cell ISO threads 3 2850 51 41D 159 001 345 Integral 2850 system Digital S3 L output...

Отзывы: