2850 Conductivity/Resistivity Sensor Electronics

Signet offers conductivity simulators in 

fi

 ve different values. These tools enable the user to validate the performance of the 

electronics independently of the electrode. This requirement is de

fi

 ned by ASTM D 1125-95 (Standard Test Methods for Electrical 

Conductivity and Resistivity of Water) which is commonly used for USP 24 applications.

3-2850.101-1  (159 001 392) 

1.0 

μ

S ±0.1%

3-2850.101-2  (159 001 393) 

2.5 

μ

S ±0.1%

3-2850.101-3  (159 001 394) 

10.0 

μ

S ±0.1%

3-2850.101-4  (159 001 395) 

18.2 M

Ω

 ±0.1%

3-2850.101-5  (159 001 396) 

10.0 M

Ω

 ±0.1%

Before using the Simulator:

Before using the simulators, disconnect all electrodes and reset the 2850 Electronics to factory calibration:
1.  Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Turn the power on, wait at least 3 seconds
2.  Press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds).
3.  When the LED goes off, release SW1. Reset is complete.

To validate Input 1:

1.  Turn the power off.
2.  Set SW3 to accept a 0.01 cell constant. Leave SW4 in the all-OPEN position.
3.  Install the simulator into the 2850 at the Input 1 conductivity electrode terminals.
4.  Connect the 2850 output terminals to the appropriate display instrument. Turn the power on.

•  The display should indicate a conductivity value within ±2% of the simulator value.
•  If the output is x10 or x100 of the correct value, recheck step 2. The 2850 is not set to accept a 0.01 cell constant.
•  If the correct value is displayed, the electronics are calibrated to ASTM D 1125 speci

fi

 cations. Remove the simulator and 

reset SW3 to the correct cell constant before reinstalling the electrode.

•  If the correct value is not displayed, contact the factory.

To validate Input 2:

1.  Turn the power off.
2.  Set SW4 to accept a 0.01 cell constant.
3.  Install the simulator into the 2850 at the Input 2 conductivity electrode terminals.
4.  Set SW3 switches to the all-OPEN position.
5.  Turn the power on and repeat step 4.

Remember

 to reset all switches to their correct position before resuming normal operation.

Cond. - Resist.

Simulator

@ 25°C

For  use with Signet 

2850 Conductivity

NIST Traceable

Electronic Certi

fi

 cation

Содержание Signet 2850

Страница 1: ...vity test solution values for simple field calibration The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the fact...

Страница 2: ...t accept any interference received including interference that may cause undesired operation Specifications Caution Warning Danger Indicates a potential hazard Failure to follow all warnings may lead...

Страница 3: ...li Cleaners Acids Bases 0 01 Cell 0 10 Cell 1 0 Cell 10 0 Cell 20 0 Cell 3 2841 3 2842 26 8 mm 1 13 in 41 9 mm 1 65 in 3 2823 203 mm 8 0 in 89 mm 3 5 in 3 2839 59 3 mm 2 33 in 73 7 mm 2 90 in 3 2840 2...

Страница 4: ...sanitary flange fittings must be installed using this option While the 2850 5X and 2850 6X electronics cannot be submerged either model will accommodate tank installation Select any electrode with a...

Страница 5: ...q Input 2 Red S L Red GND White Shield 5VDC Black S L Red GND White Shield 3 3 Analog Output 1 Analog Output 2 if applicable if applicable Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input...

Страница 6: ...W3 or SW4 Integral 2850 Systems Single Input The 2850 51 XX and 2850 52 XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single input electronic...

Страница 7: ...to 500 S 0 to 1000 S O C C C 10 to 2 M 5 to 0 S 50 to 0 S 500 to 0 S 1 000 to 0 S C O C O 0 to 2 M 0 to 10 S 0 to 100 S 0 to 1 000 S 0 to 2 000 S C O C C 2 to 0 M 10 to 0 S 100 to 0 S 1 000 to 0 S 2...

Страница 8: ...ed if both inputs are used or if Input 2 is disabled To disable Input 2 Set SW4 4 to OPEN Place each sensor into a test solution that represents the operating range of that sensor When SW1 is pressed...

Страница 9: ...e test solution is outside the 10 tolerance The electronics or the electrode is defective Use fresh test solutions and restart the calibration Replace the electronics or the electrode After completing...

Страница 10: ...1 Turn the power off 2 Set SW3 to accept a 0 01 cell constant Leave SW4 in the all OPEN position 3 Install the simulator into the 2850 at the Input 1 conductivity electrode terminals 4 Connect the 285...

Страница 11: ...11 2850 Conductivity Resistivity Sensor Electronics Notes...

Страница 12: ...tem Digital S3 L output 0 01 cell ISO threads 3 2850 51 40D 159 001 344 Integral 2850 system Digital S3 L output 0 1 cell ISO threads 3 2850 51 41D 159 001 345 Integral 2850 system Digital S3 L output...

Отзывы: