page 7 of 12

‡ SIGNET

 2550 InsertionMagmeter

Step:

Press:

To display:

1.

7.

Example illustrates setting current  output for 0 to 300 GPM = 4 - 20 mA.

The factory default value for current is :
0 to 200 GPM = 4 - 20 mA.

4.

2.

3.

5.

10.

11.

12.

Arrow

 keys advance main menu function; 

Run/Edit

 key returns to normal operation.

6.

9.

13.

14.

15.

16.

17.

Up arrow changes flashing digit.

Down arrow advances to next menu item.

Arrow keys will toggle setting.

Enter key stores final setting and returns to main menu.

Down arrow advances to next menu item.

Press enter 5 times to store the new setting and return
to main menu.

5.3.3  Current Output

The current output of the 2550 is set via three settings.  

Low current

 specifies the flow rate represented by the minimum current signal

(4 or 0 mA).  

High current

 is the flow rate represented by the maximum current (20 mA), and 

current

 selects a 4 to 20 or a 0 to 20 mA

span.  To change the current selections, first enter the access code (section 5.2.1) then press the down arrow until the display shows:

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

8.

5.4 PerformanceFunctions

The next five main menu functions help to tailor the magmeter
signals to your specific environment.

5.4.1

Sensitivity

The sensitivity function limits the amount of reference level noise
detected by the Magmeter input circuit.  Use the StdE display
(section 5.5.2) to evaluate the amount of reference noise in your
system.  If StdE is greater than 0.05, decreasing the sensitivity
value may stabilize flow readings.

The range selections are from 1 to 10, with 1 representing the
most sensitive and 10 the least sensitive.

To change the sensitivity setting, first enter the access code
(section 5.2.1) then press the down arrow until the display shows:

Example illustrates changing sensitivity setting to 6.

The factory default setting
for sensitivity  is 4.

The enter key stores the new
setting and returns the
display to the main menu.

1.

2.

3.

4.

Up arrow changes the setting.

ENTER

ENTER

Step: Press:  To display:

Arrow

 keys advance main menu function; 

Run/Edit

 key returns

to normal operation.

Содержание SIGNET 2550

Страница 1: ...es automatic temperature compensation resulting in outputs that are accurate within 2 of actual flow rate The bipolar electronic design and the 10000 MΩ input impedance reduce galvanic formation on the electrodes and minimize coating problems The isolated current output provides an universal signal to recorders valves process control and data acquisition devices The open collector frequency output...

Страница 2: ... Inlet Outlet 90 elbow Inlet Outlet Measure and record the interior diameter of the pipe This value will be required to calibrate the 2550 Interior diameter ________________________________ 3 3 Sensor Installation with tool WARNING Do not remove safety ring or adjustment nut from sensor The installation tool 3 2550 355 simplifies installation and insures proper insertion depth for maximum accuracy...

Страница 3: ...aptive outer cover screws 1 lift off outer cover 2 then loosen two captive screws 3 and remove access plate 4 A wiring diagram is located on the inside of the access plate Five terminal strips are available TS 3 Analog output 4 to 20 mA or 0 to 20 mA output terminal 1 positive R terminal 2 negative R 18 to 22 AWG wire recommended TS 4 Serial communication Future expansion not available in this mod...

Страница 4: ...oop Without moisture drip loop Careful cable routing and sealed ports can help prevent water intrusion 3 In plastic piping systems the fluid carries significant levels of static electricity that must be grounded for best magmeter performance Since plastic pipe is not grounded as metal pipes are extra steps must be taken Identify the places where the fluid is in electrical contact with external met...

Страница 5: ...played user calibration displayed This basic operating procedure repeats throughout the 2550 program 1 Press RUN EDIT to enter edit menu 2 Press to move to a specific menu item 3 Press ENTER key to select the item for editing 4 Press ENTER to edit the value selection 5 Press ENTER to store the new value selection 6 Press to select another menu item or 7 Press RUN EDIT to return to normal operation...

Страница 6: ...r diameter You must enter the precise interior diameter of your pipe as measured during the installation process Entry is made in inches when the dimension entered is less than 40 and automatically switches to millimeters when the dimension entered is greater than 40 To change the id selection enter the access code section 5 2 1 then press the down arrow until the display shows 3 2 1 4 7 6 5 8 9 1...

Страница 7: ...tion 5 2 1 then press the down arrow until the display shows ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER 8 5 4 PerformanceFunctions The next five main menu functions help to tailor the magmeter signals to your specific environment 5 4 1 Sensitivity The sensitivity function limits the amount of reference level noise detected by the Magmeter input circuit Use the St...

Страница 8: ...ow to change the setting to 2 The factory default setting for viscosity is 1 3 2 1 ENTER ENTER Step Press To display Viscosity ranges 1 1 cp water 2 2 cp 3 4 cp 4 8 cp 5 N A future expansion Step Press To display Arrow keys advance main menu function Run Edit key returns to normal operation 5 4 4 AC Noise Filter Select 50 Hz or 60 Hz rejection filter according to the AC power standard in your area...

Страница 9: ...ground Use the Sensitivity adjustment section 5 4 1 to compensate and check the grounding requirements section 3 2 The standard flow deviation StdF is an overall indication of the stability of the flow signal It incorporates the StdE as 1 2 3 RUN EDIT ENTER well as mechanical factors such as unstable flow profiles misaligned sensors and coated electrodes StdF is displayed as a percentage of flow r...

Страница 10: ...40S 8 in 7 98 155 928 590 26 9 62 6 24 2 54 23 61 steel sch 80S 8 in 7 63 142 328 538 77 10 54 5 69 2 78 21 55 PVC sched 80 10 in 9 49 220 605 835 09 6 80 8 82 1 80 33 40 PVC sched 40 10 in 9 98 243 625 922 23 6 16 9 75 1 63 36 89 steel sch 5S 10 in 10 48 268 966 1018 2 5 58 10 76 1 47 40 73 steel sch 10S 10 in 10 42 265 794 1006 1 5 64 10 63 1 49 40 24 steel sch 40S 10 in 10 02 245 779 930 38 6 1...

Страница 11: ...33 54 505 44 11 23 5 34 2 97 20 22 PVDF bar 10 250 234 8 209 19 791 78 7 17 8 37 1 89 31 67 PVDF bar 16 63 57 12 33 46 661 121 67 0 49 32 15 1 87 PVDF bar 16 75 67 8 17 44 66 019 86 00 0 70 22 72 2 64 PVDF bar 16 90 81 4 25 14 95 161 59 66 1 01 15 76 3 81 PVDF bar 16 110 99 4 37 49 141 9 40 01 1 50 10 57 5 68 PVC bar 10 63 57 12 33 46 661 121 67 0 49 32 15 1 87 PVC bar 10 75 67 8 17 44 66 019 86 0...

Страница 12: ...x 25 130 Hz 5 2 ft s 130 Hz Unit GPM Display 120 GPM 120 GPM 130 Hz A 120 130 0 92 K 60 0 92 65 2 6 Ordering Information Mfr Part No Code Description 3 2550 100 110 159 000 294 Insertion Magmeter with 2 in NPT Fitting 3 2550 100 110T 159 000 632 Insertion Magmeter with 2 in NPT Fitting and Installation Tool 3 2550 100 111 198 840 024 Insertion Magmeter with 2 in ISO Fitting 3 2550 100 111T 198 840...

Отзывы: