Maintenance and Cleaning

The 2537 requires very little maintenance.

•  If the paddlewheel becomes fouled, it can be cleaned with mild detergents and a small brush.

•  The electronics portion of the 2537 does not require maintenance or cleaning.

Rotor Replacement Procedure

1.  To remove the rotor, insert a small screwdriver between the rotor and the ear of the sensor.

2.  Twist the screwdriver blade to 

#

 ex the ear outward enough to remove one end of the rotor and pin. 

DO NOT 

#

 ex the ear any more than necessary! If it breaks, the sensor cannot be repaired.

3.  Install the new rotor by inserting one tip of the pin into the hole, then carefully 

#

 ex the opposite ear 

back enough to slip rotor into place.

H-Dimensions

The plastic insert in Weldolet 

!

 ttings MUST be removed during the welding process. When reinstalled, it is important that the insert be 

threaded to the proper height ("H" dimension).

Weldolet

“H” dimension

Part number inches

mm

CS4W020

2.38

60.45

CS4W025

2.33

59.18

CS4W030

2.32

58.92

CS4W040

2.30

58.42

CS4W050

3.09

78.48

CS4W060

2.96

75.18

CS4W080

2.73

69.34

"H"

Weldolet

“H” dimension

Part number inches 

mm

CR4W020

2.38

60.45

CR4W025

2.33

59.18

CR4W030

2.32

58.92

CR4W040

2.30

58.42

CR4W050

3.09

78.48

CR4W060

2.96

75.18

CR4W080

2.73

69.34

Содержание Signet 2537

Страница 1: ...Digital S3 L or 4 to 20 mA This unit can be con gured on site directly through the built in user interface The 4 to 20 mA model provides a blind current loop output The Digital S3 L model provides a...

Страница 2: ...re and concentration Utilizing sensors in applications with aggressive substances is possible On special request GF Signet can provide sensors with a different internal resin encapsulation potting tha...

Страница 3: ...when the system is in operation These con gurations are not recommended because it is dif cult to keep the pipe full Installation Pipe fittings Georg Fischer Signet offers a wide selection of install...

Страница 4: ...irectly into the conduit port For conduit with ISO threads use the black thread adapter included with the connector kit To comply with NEC requirements do not use any metal conduit in the installation...

Страница 5: ...e Dry Contact Relay requires an external power source of 24 VDC 10 The Solid State Relay requires an external power source of 5 to 24 VDC Solid State Relay requires a pull up resistor 10K ohm recommen...

Страница 6: ...key S1 is held down for three seconds the ow rate is displayed for 10 minutes before reverting back to the model name In the Multi Model if the Multi menu item is set to divider then the divided puls...

Страница 7: ...0 to 9 page 9 Mode Select the output mode Total or Divider or On OFF page 10 P Factor As PULSE DIVIDER divides input freq 1 0000 to 99999 page 10 P Factor As PULSE TOTAL multiplies K Factor 1 0000 to...

Страница 8: ...tion Save the new setting Select the volumetric units for the ow measurement Flow Units Example Set the Flow Units to Gallons per minute This function is available on these versions of the 2537 2s 2s...

Страница 9: ...ING only AVERAGING and SENSITIVITY Averaging and Sensitivity Settings Because ideal ow conditions are often impossible to achieve the ow rate is often erratic which causes erratic readings in control...

Страница 10: ...Factor MULTIPLIED by the P factor The resulting frequency is output through a relay 2s Go to next menu item Return to Normal Operation Save the new setting Select DIVIDER or TOTAL operation The facto...

Страница 11: ...hes the setpoint the 2537 will wait this long in seconds before triggering the alarm The factory setting is 0000 0 seconds 4 Program the DELAY A HI alarm will be activated when the ow rate rises ABOVE...

Страница 12: ...T020 54 468 14 391 STAINLESS STEEL TEES ON SCH 40 PIPE 1 2 CR4T005 734 20 193 98 3 4 CR4T007 412 10 108 88 1 CR4T010 252 70 66 764 1 1 4 CR4T012 128 12 33 849 1 1 2 CR4T015 77 320 20 428 2 CR4T020 45...

Страница 13: ...00 1 9868 8 CR4W080 4 3400 1 1466 CARBON STEEL WELDOLETS ON SCH 40 PIPE 2 1 2 CS4W025 37 600 9 9339 3 CS4W030 24 340 6 4306 4 CS4W040 13 920 3 6777 5 CS4W050 10 860 2 8692 6 CS4W060 7 5200 1 9868 8 CS...

Страница 14: ...it breaks the sensor cannot be repaired 3 Install the new rotor by inserting one tip of the pin into the hole then carefully ex the opposite ear back enough to slip rotor into place H Dimensions The p...

Страница 15: ...15 Signet 2537 Flowmeter Notes...

Страница 16: ...Digital S3 L Integral Mount 3 2537 6C T0 159 001 320 4 to 20 mA Integral Mount Accessories Mfr Part No Code Description 3 2536 320 1 198 820 052 Rotor PVDF Black 3 2536 320 2 159 000 272 Rotor PVDF Na...

Отзывы: