2610 Calibration
Calibration
settings:
1
Click to Start Dissolved Oxygen Saturation
2
Click to Reset Dissolved Oxygen Calibration to Factory Default
First Point
settings:
3
With probe in 100% Saturated Water click Start to begin the Calibration
4
Once probe reading is stable click Continue
Second Point
settings:
5
With probe in 0% Saturated Water click Start to begin the second point Calibration
6
Once probe reading is stable click Finish to complete calibration
Reset Calibration:
7
Click Reset Cal to reset the DO calibration to Factory Default
1
2
3
4
5
6
7
15
GF 0252 Configuration Tool
Содержание Signet 0252
Страница 11: ...258X Operation 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 11 GF 0252 Configuration Tool...
Страница 13: ...258X Calibration 6 7 8 9 10 11 13 GF 0252 Configuration Tool...
Страница 17: ...2751 Operation 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 17 GF 0252 Configuration Tool...
Страница 19: ...2751 Calibration 5 6 7 8 9 10 11 12 19 GF 0252 Configuration Tool...
Страница 20: ...2751 Calibration 13 14 15 16 17 20 GF 0252 Configuration Tool...
Страница 22: ...9900 Operation 2 3 5a 5b General Software Operation pg 7 22 GF 0252 Configuration Tool...