Final inspection
• If the machine has dosing of cleaning agent: Press
S
and check to make
sure the lamp in the push button lights.
• Check to make sure the incoming hot water has the correct temperature. It
should be between 45 and 60°C.
• Start the ’Normal’ program with empty machine and check that:
- The PROCESS IN OPERATION lamp is ON.
- The temperature rises continuously until its cooling phase begins.
- Any pause for heating is not longer than about 10 minutes (applies to
apparatus connected to steam and apparatus with steam generator).
- The final temperature is higher than 90°C.
- The PROCESS COMPLETE lamp lights when the program is over.
Fig. 10. Indicating lamps
L M
K
Process in operation
Process completed
• If the machine has a system for cleaning or deliming agent, or both:
Pour a sufficient amount of agent into a measuring glass. Make sure that the
suction hose is full of agent. Start a ’normal’ program and check that the
disinfector uses the correct amount. For cleaning agent, the amount per
program is normally about 30 ml, and for deliming agent about 1,5 to 4,5 ml
(follow the supplier’s recommendations).
N.B.
When taking the reading, the suction pipe should not be in the
measuring glass.
• Check that there are no leaks. Tighten if necessary.
Содержание S-406
Страница 15: ...15 Installation Flusher disinfector S 406 9904...
Страница 16: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ......
Страница 24: ...This product is manufactured by GETINGE DISINFECTIONAB Ljungadalsgatan 11 Box 1505 351 15V xj Sweden...