Malfunctions
- continued -
Fault finding list for troubleshooting
- continued -
If faults occur that are not listed above or cannot be corrected, please contact our service centre
or authorized agency in your country.
Equipment does not work correctly
Fault:
Indicated conductivity reading is larger than the reference value.
Remedy:
Change correction factor (3. Correction of measured value) and, if necessary,
temperature coefficient (4.1 TK).
Fault:
Indicated conductivity reading is smaller than the reference value.
Remedy:
Change correction factor (3. Correction of measured value) and, if necessary,
temperature coefficient (4.1 TK).
Fault:
The measured value cannot be adapted by changing the correction factor or the
temperature coefficient.
Remedy:
Remove conductivity electrode and clean measuring surface.
Fault:
Current output is always 20 mA.
Remedy:
Adapt parameter “6. Standardisation of current output” to the conductivity reading.