                             Operating Instructions Centrifuge Micro III 

 

 

Gerber Instruments AG                                     3                                 Version 1.2 (03/2021) 

 

Warning symbol: Attention, possible dangerous situation 

Warning symbol: Dangerous electrical voltage 

1. 

Introduction 

 

1.1 

Use 

 
The small centrifuge Micro III is used for the determination of fat in milk and dairy products 
according to Gerber or Babcock.

 

 
 

1.2 

Description 

 
The centrifuge Micro III has high safety standards and reliability and following  
product features: 
 

  Strong sheet steel housing with an acid-resistant paint work 
  Cover lock to prevent opening of the centrifuge during operation 
  Motor with elastic bearing for a smooth run 
  NEW: Increased precision thanks to the controlled speed 
  NEW: Preheating option 
  NEW: Unbalance switch-off in case of uneven filling or sudden glass breakage 

 
 

1.3 

Ambient conditions 

 

  Permissible ambient tempe10 °C to +40 °C 
  Max. humidity 80 % 
  Lowest storage and transport temperature -20 °C 

 
 

2. 

Safety instructions 

 

2.1 

Symbols of the operating instructions 

 
 

   

 
 

 

 
 

   

 
 
 
 
 
 
 
 
 
 

 

Содержание Micro 3

Страница 1: ...Centrifuge Micro III OPERATING INSTRUCTIONS ...

Страница 2: ...on of centrifuge Micro III 6 4 Unpacking Device control 7 4 1 Unpacking 7 4 2 Device control 7 4 3 Control accessories 7 4 4 Avoiding damages on the device 7 5 Description of operating unit 8 5 1 Segment display 8 5 2 LED displays 8 5 3 Button functions 9 6 Commissioning 10 6 1 Initial start 10 6 2 Setting the preheating function 10 6 3 Setting time and temperature 11 7 Daily work 11 8 Cleaning an...

Страница 3: ...icro III Gerber Instruments AG 2 Version 1 2 03 2021 11 Packaging 14 12 Disposal 14 13 Technical service 14 14 Warranty 15 15 Order numbers centrifuge Micro III 16 16 Order numbers accessories Micro III 16 17 Technical data Micro III 17 ...

Страница 4: ...gh safety standards and reliability and following product features Strong sheet steel housing with an acid resistant paint work Cover lock to prevent opening of the centrifuge during operation Motor with elastic bearing for a smooth run NEW Increased precision thanks to the controlled speed NEW Preheating option NEW Unbalance switch off in case of uneven filling or sudden glass breakage 1 3 Ambien...

Страница 5: ...igue must not operate the device Users must be familiar with the contents of the operating instructions Do not use the device with a damaged cable or plug The use of accessories which are not original spare parts from Gerber Instruments results in loss of warranty Switch off disconnect the device from the mains if it is not being used if parts are replaced and if the device is being cleaned If you...

Страница 6: ... The centrifuge can be stopped at any time Press the START STOP button 2 5 Safety instructions for cover lock The centrifuge may only be operated with the cover closed A cover lock blocks the cover closure during the run and allows you to open only at standstill In case of cut off of electricity the lid may be opened manually see 10 2 2 6 Safety instructions for unbalance switch off The centrifuge...

Страница 7: ...uctions Centrifuge Micro III Gerber Instruments AG 6 Version 1 2 03 2021 3 Description of centrifuge Micro III Cover Housing Main switch and mains plug connection Rotor with inserts Operating unit 1 2 3 4 5 1 2 5 4 3 ...

Страница 8: ...on damages Other irregularities If one of these criteria applies immediately inform the forwarding agent or the supplier 4 3 Control accessories Check with the delivery note and your order if the scope of delivery is complete Type of centrifuge Sleeves Specific country cable Fork key Wire for emergency unlocking Operating instructions 4 4 Avoiding damages on the device Do not drop or disassemble t...

Страница 9: ...ual temperature in the operating mode centrifugation Factory setting 55 C as the reaction milk sulphuric acid also generates heat 5 2 LED displays LED Brake This LED is simple and lights up red shortly before the end of the braking process LED Time This LED is dual In standby mode it lights up red In the operating mode centrifugation it lights up green and starts flashing LED Rotor This LED is dua...

Страница 10: ...3 2021 5 3 Button functions Button SET Setting temperature and time Setup mode Button Increase of temperature and time Button Reduction of temperature and time Button START STOP Start the centrifuge stops the centrifuge before the set time has elapsed Pre heating function 8 9 10 11 ...

Страница 11: ...er supply Turn on the green main switch on the right side A short click sounds the lid is unlocked The LED Time and Rotor light up red The segment displays show the factory set time 5 min and temperature 55 C Open the cover The time segment display shows LId OPEn Remove the foam from the rotor Now insert the sleeves Generally all the places of the rotor should be filled with sleeves 6 2 Setting th...

Страница 12: ...eves To avoid unbalance the opposite places in the rotor must always be filled The butyrometers must be nearly identical in terms of weight Close the cover Set the desired temperature and duration Press the button Start Stop the safety interlock is activated The LED Rotor lights up green the LED Heating lights up red The current temperature is displayed the rotor accelerates Once the rotor has rea...

Страница 13: ...pletely If sulphuric acid has leaked out clean the interior of the centrifuge and the sleeves immediately in accordance with the guidelines for handling acids 9 Checking the unbalance switch off The function of the unbalance switch off should be checked at regular intervals This work may only be carried out by authorized personal The centrifuge must be supervised during this check Check that all s...

Страница 14: ...ions for cover locking Centrifuge starts but stops immediately again Electronic defect Contact the service of Gerber Instruments AG Ibal Unbalance in the centrifuge Press the stop button open the cover and remove the causes of the unbalance Acoustic alarm sounds and other failure indications Make a note of the error message make a note of the S N Contact the service of Gerber Instruments AG 10 2 E...

Страница 15: ...f its life time the instrument must be disposed or collected in the appropriate ways according to the regulations in force in the country of use For further information please contact our sales office Email info gerber instruments ch Phone 41 0 52 343 37 37 Fax 41 0 52 343 30 70 13 Technical service Address Gerber Instruments AG Im Langhag 12 CH 8307 Effretikon Phone 41 52 343 37 37 Fax 41 52 343 ...

Страница 16: ...ce has to be returned during the period of guarantee The product needs to be returned by the customer to the company Gerber Instruments AG carriage forward Guarantees do neither effect a prolongation of the period of guarantee nor a claim on a new guarantee Any repairs due to usual wear will not be performed free of charge The customer pays the freight cost for replacement parts Gerber Instruments...

Страница 17: ...AC 50 60 Hz 16 Order numbers accessories Micro III Order No Description 03 17000 Operating and control unit 03 17001 Frequency converters 03 17002 Unbalance sensor 03 17003 Heating 03 17004 Main switch 03 17005 Temperature sensor 03 17006 Motor 03 17007 Relay for heating 03 17008 Power supply 03 17009 Cover locking 03 17010 Sensor cover locking 03 16216 Gerber rotor for 8 samples 03 16217 Gerber r...

Страница 18: ...230 VAC Frequency 60 or 50 Hz Power heating 900 W Speed 1280 rpm 50 rpm fixed program Centrifugal power 350 x g permanent adjusted Temperature Thermostatically controlled 5 C above ambient to 65 C factory set at 55 C Dimensions WxDxH 430x530 x 288 mm Weight netto 23 5 kg without sleeves Weight total up to 28 5 kg incl accessories and packaging Warranty 2 years Test certificate CE ...

Отзывы: