background image

 
 
 
 

 

Spirostik 

User Manual 

Version 1.2.2 

 

Date 24th October, 2014 

Содержание Spirostik

Страница 1: ...Spirostik User Manual Version 1 2 2 Date 24th October 2014...

Страница 2: ...ectromagnetic Compatibility 7 2 2 2 Safe Use of Spirostik 8 2 2 2 1 Connecting Spirostik double hose and handle 9 2 2 2 2 Replace handle 10 2 2 2 3 Connecting the Flow sensor into the handle 11 2 2 2...

Страница 3: ...integrated into even the largest of requirements This document contains important information for the operation of Blue Cherry and Spirostik We strongly recommend reading this manual carefully in ord...

Страница 4: ...y software has a modular design to allow the addition of further options in the future such as MVV or Pulse Oximetry measurements The main application of Spirostik is in the diagnosis of pulmonary dis...

Страница 5: ...ncludes the following components Component Name and Description USB Spirostik Electronic module connecting to USB port Pressure tube Connects flow sensor to pressure transducer Single use flow sensors...

Страница 6: ...ains instructions on how to use the software and how to handle the device but must not be seen as a training course for technicians or engineers Before using the device all users should read this manu...

Страница 7: ...EN 60601 1 Medical electrical device DIN EN 60601 1 1 General safety requirements DIN EN 60601 1 2 EMV DIN EN 60601 1 4 Programmable electrical medical devices DIN EN ISO 14971 Risk management DIN EN...

Страница 8: ...atient and non medical computer systems Flow sensors and pressure tubes must not be cleaned Humidity inside the pressure tube may affect the accuracy of the measurement Replace the pressure tube in ca...

Страница 9: ...r to ensure the continued function of the equipment this connection should be examined once per day Please read the following carefully The double hose should be securely connected to the sealing rubb...

Страница 10: ...n of the Spirostik the handle needs to be changed after not more than 12 months On the back plate of the Spirostik device there are 2 connectors where the handle tubing can be directly connected Make...

Страница 11: ...e the flow sensor is straight and the two guidance ribs are either side of the connector with a firm pressure push the flow sensor into the handle this should snap into place with the connector holes...

Страница 12: ...handle In order to protect the user from cross contamination the flow sensor can be removed from the handle with a single hand without touching the used flow sensor Please read the following procedur...

Страница 13: ...OSTIK_MAN_ENG_V1 2 2_REV01 DOCX Page 13 Gently pushing the upper part of the handle with the thumb opens the mouth of the handle so that the flow sensor can fall out This procedure can be done directl...

Страница 14: ...and cables are connected properly when using the measurement device together with other devices All connected devices must comply with legal regulations and the total leakage current must not exceed...

Страница 15: ...A within patient area and resource B in medical used room Resource A within patient area and resource B in medical used room See remarks below table Resource A within patient area and resource B in a...

Страница 16: ...t in accordance with local directives 2 2 5 Product labelling Example of a typical product labelling SN designates the serial number for the device The first 2 digits define the manufacturing year The...

Страница 17: ...o and electronic devices IPX0 Indicates no protection against dripping water Manufactured according RoHS directive 2002 95 EC 0494 Geratherm Respiratory devices carry this CE no according to directive...

Страница 18: ...be provided by Geratherm Respiratory or an authorized service partner of Geratherm Respiratory Geratherm Respiratory will not accept warranty claims on any unauthorised repairs 3 2 Warranty exclusion...

Страница 19: ...esolution Sample rate Accuracy GR Spiraflow Differential pressure 16l s 0 09kPa l s 15l s 20ml 1ml s 125 Hz 3 or 20ml s Volume Volume Accuracy 0 20L 3 or 50mL Operating conditions Temperature Air humi...