LCD shows 
“Er 0”

Pressure system is 
unstable before meas-
urement.

Don’t move 
and try again.

LCD shows 
“Er 1”

Systolic pressure 
detect fail

LCD shows 
“Er 2”

Diastolic pressure 
detect fail

LCD shows 
“Er 3”

Pneumatic system 
blocked or cuff is too 
tight during in

fl

 ation

Place the cuff 
correctly and 
try again

LCD shows 
“Er 4”

Pneumatic system 
leakage or cuff is too 
loose during in

fl

 ation

LCD shows 
“Er 5”

In

fl

 ation above 

300 mm Hg

Measure 
again after 
5 minutes. If 
the monitor is 
still abnormal, 
please contact 
the local 
distributor or 
the factory.

LCD shows 
“Er 6”

Pressure is over 
15 mm Hg for longer 
than 3 min. At the start 
of measurement the 
pressure reading must 
be below 15 mm Hg.

LCD shows 
“Er 7”

EEPROM accessing 
error

LCD shows 
“Er 8”

Device parameter 
check error

LCD shows 
“Er A”

Span error

BPM no 
response when 
you press 
button. BPM 
hang

Incorrect operation or 
strong electromagnetic 
interference

Take out bat-
teries for 

fi

 ve 

minutes, and 
then reinstall 
all batteries.

Error messages

Technical alarm:

Should the recorded blood pressure (systolic or diastolic) 
lies outside the limits speci

fi

 ed in “Speci

fi

 cations” the tech-

nical alarm will appear on the display indicating either “HI” 
or “Lo”. You should seek medical assistance and check the 
accuracy of your procedure.
The limit values for the technical alarm are factory set and 
cannot be adjusted or deactivated.
The signal shown on the display will disappear automati-
cally after about 8 seconds.

Содержание GT-5907

Страница 1: ...GB Digital Blood Pressure Monitor for the upper arm INSTRUCTION MANUAL GT 5907...

Страница 2: ...inches Designated use 28 Precautionary measures 29 Warranty 31 Information you should know before operating the unit 31 About the unit Function description 32 Explanation of displays 33 Preparation f...

Страница 3: ...down comfortably with your elbows placed on the table and both feet on the ground Please do not interlock your legs during the measurement Wrap the cuff snug around your upper arm and lay your arm on...

Страница 4: ...esult in circulatory disor ders which can cause unpleasant sensations such as lo calised bleeding under the skin or temporary numbness in your arm These symptoms do not usually last long How ever if y...

Страница 5: ...ressure level What about low blood pressure In general a lower blood pressure reading is better unless causes unpleasant symptoms such as fainting and or light headedness The warranty for this blood p...

Страница 6: ...red automatically and non invasively which helps to ease the operation The Geratherm med control have 2 x 99 memories with date and time display a big LCD screen and arrhythmia detection The RCC radio...

Страница 7: ...to fit arm sizes between 22 and 30cm 8 7 to 11 81 inches Explanation of displays access for power supply cuff On Off Button LCD Display Memory Date Time Systo lic Pulse Diasto lic Low Batterie Memory...

Страница 8: ...e type of alkaline bat teries be used to avoid incompatibility Rechargeable batteries with a voltage of 1 2V are not suitable for this product Keep batteries away from small children Do not throw batt...

Страница 9: ...or use AC adapter accessory Use the device only with a medical approved stabi lized AC adapter 6V DC 600mA Note 1 No batteries are required during operation using the AC adapter 2 If AC adapter power...

Страница 10: ...ata received the machine enters Clock Mode and keeps the time and date un changed When the machine is in the RCC Mode you can skip RCC receiving and enter the Clock Mode by pressing START button b For...

Страница 11: ...l clock adjusting if no RCC signal or weak signal a When the machine is in Clock Mode press the START and MEM button at the same time abt 2 sec the machine enters Forced RCC Receiving mode then press...

Страница 12: ...the heartbeat during the pressure are falling down f After measurement the result of systolic diastolic and pulse rate will all display on LCD And irregu lar heartbeat symbol and blood pressure class...

Страница 13: ...ssment diastolic pressure mmHg systolic pressure mmHg Severe Hypertension Grade 3 Moderate Hypertension Grade 2 Mild Hypertension Grade 1 Optimal Normal High Normal Blood pressure classification Systo...

Страница 14: ...ress the MEM but ton previously memories will show by turns d All memories have the WHO classification and the COL OUR CHANGE LED will also indicate the blood pres sure classification e If it has no m...

Страница 15: ...o not drop the unit It is not shock proof Do not modify or disassem ble the unit or the arm cuff Do not twist the arm cuff Use a cloth moistened with water or neutral detergent to clean the body of th...

Страница 16: ...not be used for a long time Do not press the START button if the cuff has not been properly wrapped around the arm Care and maintenance It is recommended the cuff should be disinfected 2 times every w...

Страница 17: ...serious arrhyth mia to use this blood pressure monitor LCD shows low battery symbol Batteries low Change the batteries Observe the applicable regulations when disposing of the device and batteries Thi...

Страница 18: ...of measurement the pressure reading must be below 15 mm Hg LCD shows Er 7 EEPROM accessing error LCD shows Er 8 Device parameter check error LCD shows Er A Span error BPM no response when you press bu...

Страница 19: ...80 pulse beats min Accuracy 3mmHg blood pressure 5 pulse rate Memory 2 x 99 times with time and date displaying Low Battery Indicator yes Auto Power Off after 1 minute Battery Life Approx 200 measurem...

Страница 20: ...ometers part 3 Supplementary requirements for electromechanical blood pressure measuring systems EN 1060 4 Non invasive sphygmomanometers Part 4 Test procedures to determine the overall system accu ra...

Страница 21: ...requirements for EMC electromagnetic compatibility and defines the levels of immunity to electromagnetic interference and the maximum electromagnetic emission levels for medical devices This blood pr...

Страница 22: ...umidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels character istic of a typical location in a typicalco...

Страница 23: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity ofequipment marked with the following symbol...

Отзывы: