Guidance and manufacturer´s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environments listed below, and 
should only beused in such environments:

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1 

RF energy is used only to maintain 
device’s operation. Therefore, its RF 
emissions are so low that it’s not likely to 
cause any interference in nearby elec-
tronic equipment.

RF emissions CISPR 11 

Class B 

The device is suitable for use in all 
establishments, including domesticestab-
lishments, and those directly connected 
to the public low-voltage power supply 
networkthat supplies buildings used for 
domestic purposes.

Harmonic  emissions                                             
IEC 61000-3-2 

Class A 

Voltage 

fl

 uctuations / 

fl

 icker 

emissions IEC 61000-3-3 

complies 

Guidance and manufacturer´s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environments listed below, and 
should only beused in such environments:

Immunity test

IEC 60601 – 
test level

Compliance 
level

Electromagnetic environment 
- guidance

Electrostatic 
discharge (ESD) 
IEC 61000-4-2

± 6 kV contact 
± 8 kV air 

± 6 kV contact 
± 8 kV air 

Floors should be wood, concrete 
or ceramic tile. If 

fl

 oors are 

covered with synthetic material, 
the relative humidity should be at 
least 30 %.

Power frequency
(50/60 Hz) 
magnetic 

fi

 eld    

IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic 

fi

 elds 

should be at levels character-
istic of a typical location in a 
typicalcommercial or hospital 
environment.

Recommended separation distances between portable and mobile RF communication

The device is intended for use in an electromagnetic environment where radiated RF 
disturbancesare under control. User can help prevent electromagnetic interference by 
keeping the device at a minimum distance from portable and mobile RF communications 
equipment (transmitters). Below table details the maximum output power of transmitter:

Rated maximum 
output power 
of transmitters 
in Watt

Seperation distance according to frequency of transmitter / m

150 kHz to 80 MHz
d = 1.2 

P

80 MHz to 800 MHz
d = 1.2 

P

800 MHz to 2.5 GHz 
d = 2.3 

P

0.01
0.1
1
10
100

0.12 
0.38 
1.2 
3.8 
12

0.12 
0.38 
1.2 
3.8 
12

0.23 
0.73 
2.3 
7.3 
23

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in metres (m) can be estimated using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.

NOTE 1:

At 80 MHz and 800 MHz, the separation distance for the higher fre-
quency range applies.

NOTE 2:

These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and re

fl

 ection from structures, objects 

and people. 

Содержание GT-5907

Страница 1: ...GB Digital Blood Pressure Monitor for the upper arm INSTRUCTION MANUAL GT 5907...

Страница 2: ...inches Designated use 28 Precautionary measures 29 Warranty 31 Information you should know before operating the unit 31 About the unit Function description 32 Explanation of displays 33 Preparation f...

Страница 3: ...down comfortably with your elbows placed on the table and both feet on the ground Please do not interlock your legs during the measurement Wrap the cuff snug around your upper arm and lay your arm on...

Страница 4: ...esult in circulatory disor ders which can cause unpleasant sensations such as lo calised bleeding under the skin or temporary numbness in your arm These symptoms do not usually last long How ever if y...

Страница 5: ...ressure level What about low blood pressure In general a lower blood pressure reading is better unless causes unpleasant symptoms such as fainting and or light headedness The warranty for this blood p...

Страница 6: ...red automatically and non invasively which helps to ease the operation The Geratherm med control have 2 x 99 memories with date and time display a big LCD screen and arrhythmia detection The RCC radio...

Страница 7: ...to fit arm sizes between 22 and 30cm 8 7 to 11 81 inches Explanation of displays access for power supply cuff On Off Button LCD Display Memory Date Time Systo lic Pulse Diasto lic Low Batterie Memory...

Страница 8: ...e type of alkaline bat teries be used to avoid incompatibility Rechargeable batteries with a voltage of 1 2V are not suitable for this product Keep batteries away from small children Do not throw batt...

Страница 9: ...or use AC adapter accessory Use the device only with a medical approved stabi lized AC adapter 6V DC 600mA Note 1 No batteries are required during operation using the AC adapter 2 If AC adapter power...

Страница 10: ...ata received the machine enters Clock Mode and keeps the time and date un changed When the machine is in the RCC Mode you can skip RCC receiving and enter the Clock Mode by pressing START button b For...

Страница 11: ...l clock adjusting if no RCC signal or weak signal a When the machine is in Clock Mode press the START and MEM button at the same time abt 2 sec the machine enters Forced RCC Receiving mode then press...

Страница 12: ...the heartbeat during the pressure are falling down f After measurement the result of systolic diastolic and pulse rate will all display on LCD And irregu lar heartbeat symbol and blood pressure class...

Страница 13: ...ssment diastolic pressure mmHg systolic pressure mmHg Severe Hypertension Grade 3 Moderate Hypertension Grade 2 Mild Hypertension Grade 1 Optimal Normal High Normal Blood pressure classification Systo...

Страница 14: ...ress the MEM but ton previously memories will show by turns d All memories have the WHO classification and the COL OUR CHANGE LED will also indicate the blood pres sure classification e If it has no m...

Страница 15: ...o not drop the unit It is not shock proof Do not modify or disassem ble the unit or the arm cuff Do not twist the arm cuff Use a cloth moistened with water or neutral detergent to clean the body of th...

Страница 16: ...not be used for a long time Do not press the START button if the cuff has not been properly wrapped around the arm Care and maintenance It is recommended the cuff should be disinfected 2 times every w...

Страница 17: ...serious arrhyth mia to use this blood pressure monitor LCD shows low battery symbol Batteries low Change the batteries Observe the applicable regulations when disposing of the device and batteries Thi...

Страница 18: ...of measurement the pressure reading must be below 15 mm Hg LCD shows Er 7 EEPROM accessing error LCD shows Er 8 Device parameter check error LCD shows Er A Span error BPM no response when you press bu...

Страница 19: ...80 pulse beats min Accuracy 3mmHg blood pressure 5 pulse rate Memory 2 x 99 times with time and date displaying Low Battery Indicator yes Auto Power Off after 1 minute Battery Life Approx 200 measurem...

Страница 20: ...ometers part 3 Supplementary requirements for electromechanical blood pressure measuring systems EN 1060 4 Non invasive sphygmomanometers Part 4 Test procedures to determine the overall system accu ra...

Страница 21: ...requirements for EMC electromagnetic compatibility and defines the levels of immunity to electromagnetic interference and the maximum electromagnetic emission levels for medical devices This blood pr...

Страница 22: ...umidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels character istic of a typical location in a typicalco...

Страница 23: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity ofequipment marked with the following symbol...

Отзывы: