Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Version: 6 | Release date: 02 August 2022
Page 59
7
Operating Instructions
7.1
Checking for Worn Parts
The Ergostik Blueflow should be checked for defective wearing
parts each day before beginning treatments.
You can get a description in:
•
Chap. 8 “Servicing / Maintenance”
And:
Possible danger to life.
Reason: Misdiagnosis caused by measurement error
due to incorrect components or improper use.
Therefore:
•
Replace the flow sensor and the mouthpiece in
case of a functional error!
•
Always follow instructions (see chap. 8 "Servicing
/ Maintenance" and chap. 9 "Cleaning and
Disinfection")!
7.1.1
Minifilter / Permapure Check
The permapure drying tube and
minifilter must be checked regularly
for discoloration.
Please note that the permapure tube
and minifilter are transparent as
shown in Fig. [1] and are not
severely discoloured as shown in
Fig. [2].
If the permapure tube turns brown
during the course of use, it should
be replaced together with the
minifilter.
1
2