Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Page 72
Version: 6 | Release date: 02 August 2022
9
Cleaning and Disinfection
The devices of the manufacturer were designed in such a way
that minimal effort is required for cleaning and disinfection.
This is why just a few tasks are necessary to keep Ergostik
Blueflow functional and clean.
Cleaning and disinfection may cause discolouration of the
components, but without impairing their function.
The following intervals apply:
Component
Interval
Method
Noseclip
After each patient
Dispose of!
Blueflow Disposable
After each patient
Dispose of!
Exercise Mask Adapter
Blueflow
After each patient
Dispose of!
Blueflow, Disposable
mouthpiece
After each patient
Dispose of!
Face mask
After each patient
Cleaning and
disinfection bath
Mask net
After each patient
Cleaning and
disinfection bath
Ergostik Blueflow
Weekly
Wipe / surface
disinfection
Blueclip
After each patient
Wipe / surface
disinfection
All other touchable parts
Weekly
Wipe disinfection
Possible physical injury.
Reason: misinterpretation of obvious measuring errors
(drift) caused by liquid in the tubes. Therefore:
•
Clean tubes only externally!
•
Replace dirty tubes!