Instructions for Use
Spirostik
Serial no.: xx|8|001|yyy and 2001xxxxx
Version: 6 | Release date: 14. July 2021
Page 37
Reason: Electric shock. Misdiagnosis. Therefore:
•
Ensure that in the patient environment
−
a
distance of 1.5 meters to the patient
−
there are
no accessible electrical parts (interfaces, plugs,
etc.) that are not isolated from the mains with an
isolation voltage of 4 kV!
•
When connecting Spirostik to other devices,
make sure that all connections and cables are
secured and cannot come loose by themselves!
•
Do not connect any portable multiple sockets or
extension cables to the system!
•
Do not connect any additional USB devices
(except mouse, keyboard and printer)!
•
Do not install any further software!
Ask your authorised specialist retail partner
which Devices are approved by the
manufacturer!
Reason: Electrical shock and / or misdiagnosis due to
loss of electrical safety caused by exceeding the
recommended maintenance schedule. Therefore:
•
Regularly check the specified maintenance
schedules!
•
If a maintenance schedule is exceeded, do not
continue to use the total system! Request
maintenance work from your authorised
specialist retail partner!
Attention
Spirostik could be damaged.
Reason: Electrostatic discharges. Therefore:
•
Preferably no floor made of synthetic material!
•
Otherwise a relative air humidity of at least 30 % is
required!