Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Page 64
Version: 8 | Release date: 16. August 2021
7.2
Switching Diffustik / Variant PFTstik On / Off
When using a cart, this is done via the On / Off switch on the
cart. Otherwise the On / Off switch on the back of the Diffustik
device must be activated. The Diffustik resp. the variant PFTstik
is designed for permanent operation and doesn't need to be
switched on and off daily.
For descriptions of "Recommissioning after Servicing and
Cleaning Work" see:
•
Chap. 6.2 "Recommissioning after Servicing / Cleaning
Work"
7.3
Calibrating Diffustik / Variant PFTstik
There are various processes to be carried out for Diffustik resp.
the variant PFTstik in order to calibrate or validate the volume
measurement.
The required intervals can be found in the tables below.
Interval
Calibration
Once per day, and after changing the flow
sensor
Volume calibration
Once per week
Flow linearity test
Once per year
Mouth Pressure Calibration
For details on individual calibrations, please read:
•
The separate IFU Calibration