Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Page 54
Version: 8 | Release date: 16. August 2021
6
Operation
For the Diffustik variant resp. when the "CO Diffusion" option is
added to the PFTstik variant, the safety instructions in the
separate IFU Add-on CO Diffusion also apply:
6.1
Initial Operating
The first commissioning of the Diffustik resp. the variant PFTstik
and its recommissioning after maintenance work in accordance
with the responsible organisation’s obligations
(see chap. 8 "Servicing / Maintenance"), which is carried out by
qualified personnel of the specialist retail partner, is also only
carried out by these qualified personnel.
Diffustik resp. the variant PFTstik is only completely ready to
function after calibration and once the initial operation is
complete.
6.2
Recommissioning after Servicing / Cleaning Work
The responsible organisation is responsible for returning the
device to operation after servicing / cleaning work for which the
user / operator is authorised.
Servicing work here also includes the necessary checks which
are necessary if Diffustik resp. the variant PFTstik should fall
down.
Attention
Diffustik resp. the variant PFTstik could be damaged.
Reason: Careless operation. Therefore:
•
Assure that all cables and tubes are connected carefully!
•
Generally, do not expose cables, tubes and their
connections to mechanical stress such as tension,
pressure, bending or similar!