background image

TURBO INSTRUCTIONS

Seametrics • 253.872.0284 

Page16 seametrics.com

Power On Function

In order to save power, the power to the 

sensing probe on the Turbo is normally 

off,  even  when  the  Turbo  sensor  itself 

is powered. The sensing probe requires 

a warm-up or stabilization time before 

returning valid readings. The sensing probe 

is turned on when a reading is requested 

or when a Power On command is written 

to the Turbo. 

(Note: If you are recording sessions, reading 

with Aqua4Plus, or using SDI-12, you do not 

need to be concerned with the stabilization 

period. It is automatically taken care of 

in these situations. You only need to be 

concerned when using your own device to 

take Modbus readings.)

There is a “power on” register on the 

Turbo (register 62720 or 0xF500).  When a  

positive value “n” is written to this register, 

the power is turned on to the sensing 

probe and remains on for n/4 seconds. 

When a reading is requested, the timer 

is reset and the sensing element remains 

on for another n/4 seconds. To force the 

power off immediately after a reading, for 

power savings, write a zero to the register. 

If the power to the entire Turbo sensor is 

turned off, this value defaults to 48 (or 12 

seconds).
The recommended procedure to ensure 

proper warm-up/stabilization and to 

conserve power is:

•  Write Power On value of  48 (48/4 = 

12 seconds)

•  Power turns on

•  Starts sampling about two times 

per second

•  Wait 10 seconds then request a 

reading

•  Returns most recent sample

•  Resets Power On value to 48 

(48/4 = 12 seconds)

•  Write Power On value of 0 (zero)

• 

Turns power off

•  Retains last reading

If you Cannot Write to the Power On 

Register

If you are reading your device using a meter 

or other device that cannot write the Power 

On but simply takes readings on a specified 

schedule, be sure to set the polling interval 

to less than 12 seconds. This will ensure 

that the sensing probe is always powered 

up and readings should be fresh and stable. 

Note that the first reading when you turn 

on the sensor and meter will be old but will 

refresh within a second or two.

 For further information and detailed 

Modbus examples, refer to our website at 

www.seametrics.com.

Data Format

The data is returned as 32-bit IEEE floating-

point values, high word first, also referred 

to as big-endian or float inverse.

Wiper

Before each reading, the sensor will apply 

power to the turbidity sensing unit and 

run the wiper once. If your polling interval 

is 10 seconds or less, the sensing unit will  

remain on and will only wipe once when 

first started.

APPENDIX C: READING THE SENSOR

Содержание Seametrics Turbo

Страница 1: ...mpa n y ISO 9001 Precision Environmental Sensors An ONICON Brand Turbo Tubidity Smart Sensor with Data Logging For Sales Service Contact 2650 E 40th Ave Denver CO 80205 Phone 303 320 4764 Fax 303 322...

Страница 2: ...TURBO INSTRUCTIONS Seametrics 253 872 0284 Page2 seametrics com 1997 2016 Seametrics All rights reserved Registered trademarks and trademarks belong to their respective owners...

Страница 3: ...uter 5 Installing the Aqua4Plus Software 5 Installing the Sensor 5 Real time Data 6 Data Logging 6 Reports 8 Maintenance 11 Erratic Readings 11 Grounding Issues 11 Appendix A Technical Specifications...

Страница 4: ...signs of damage on the appropriate shipping form After opening the carton look for concealed damage such as a cut cable If concealed damage is found immediately file a claim with the carrier Do s and...

Страница 5: ...ing you to do the following Read via the Modbus protocol using your own software Read via SDI 12 protocol Display readings on a panel meter If you want to use one of these methods see Appendix C or co...

Страница 6: ...Start readings default to table view To switch to Real time graphing view click the graph icon Real time readings default to a 1 second interval for 1 minute to adjust enter your desired settings her...

Страница 7: ...on active under the Data Files tab as well as in the upper menu Once files have been started logged on the sensor they will be displayed under the Data Files tab Set Up Logging Window Here you will na...

Страница 8: ...lder on your PC as well See Program Settings for default data folder location In the main view you ll see a list of reports sorted by date size or file name as selected here You can also search report...

Страница 9: ...TION AND OPERATION Click to view data as a table Click to view data statistics Click to switch to full screen graphing view Graph saving and export options are available here You may change the displa...

Страница 10: ...bidity is always in NTU The Information tab is a new feature allowing users to add metadata to their reports such as site location field notes or comments The Schedule tab will display the logging set...

Страница 11: ...in the earth between the grounding points current can flow between the points creating an unexpected voltage difference and resulting erroneous readings The single most important step in minimizing a...

Страница 12: ...feet Input Power 12 Vdc nominal 9 15 Vdc Ranges 0 400 NTU 0 3000 NTU For Modbus with 5 pin connector For SDI 12 with 5 pin connector For SDI 12 without connector White Purple Yellow Brown Blue Shield...

Страница 13: ...e results Make sure the bottom face of the turbidity probe is submerged about 0 5 12mm into the calibration standard and at least 2 50mm from the bottom of the container to avoid interference with the...

Страница 14: ...r stability while Aqua4Plus is measuring to ensure an accurate calibration To accept the second point reading click Ok Your new offset and slope values will appear to the right confirm other settings...

Страница 15: ...or SDI 12 you must have version 2 2 or higher Reading Via Modbus Setting Baud Rate Your sensor comes configured to communicate at 38 400 baud with 8 data bits one stop bit and no parity The sensor can...

Страница 16: ...aults to 48 or 12 seconds The recommended procedure to ensure proper warm up stabilization and to conserve power is Write Power On value of 48 48 4 12 seconds Power turns on Starts sampling about two...

Страница 17: ...ss Query a a cr lf a cr lf Change Address a A b b cr lf change address from a to b Request measurement a M a 0022 cr lf request all measurements a D0 a 24 2412 458 73 cr lf read temperature and turbid...

Страница 18: ...f request measurements a D0 a 458 73 cr lf read turbidity Concurrent measurement with CRC a CC a 0022 cr lf request all measurements w CRC a D0 a 24 2412 458 73 crc cr lf read temperature and turbidit...

Страница 19: ...which cause damage to the products or cause abnormally frequent service calls or service problems h A new warranty period shall not be established for repaired or replaced material products or suppli...

Страница 20: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 14418r10 20190117 1 17 2019...

Отзывы: